If you have selected H/O
Select units of measure- inlb, ftlb, etc. using
Note: High Output devices usually produce an output voltage of approximately
Type in your new span; press | Enter |
Type in Serial Number; press | Enter |
Screen will indicate ‘End of
Connect your high output device and select measure mode
(refer to instruction sheet for measure mode only)
Use an independent measuring device to apply a torque to the transducer and check the read- ing on the independent unit against the reading shown on the ETA. It will be necessary to return to the ‘Edit Tx’ menu option and adjust the span setting until the two readings agree. You may have to repeat the above procedure several times until your high output transducer is correctly calibrated.
If you have selected KPD
Select location of decimal point using | then press Enter |
|
Select Parity- Odd, Even or Disabled using | then press | Enter |
To | 1 |
|
Select another memory (B to I) using the | keys and press | Enter |
To return to main menu press MENU |
|
|