Ingersoll-Rand ETA2, ETA5 manual OP Code ERR

Models: ETA5 ETA2

1 66
Download 66 pages 63.7 Kb
Page 61
Image 61

7529_eng.fm Page 60 Wednesday, November 12, 2003 8:57 AM

Code

Message

Explanation

803

OP CODE ERR

The processor has attempted to execute an

 

 

illegal instruction.

804

STACK ERR

The processor stack has overflowed.

901

TX SPEC

Illegal transducer specification number received

 

 

from PC.

902

TX TYPE

Illegal transducer type received from PC.

904

TX UNITS

Illegal transducer units data received from PC.

905

SERIAL

Illegal serial transducer data received from PC.

906

UNITS

Illegal units of measure data received from PC.

907

DIRECT

Illegal direction data received from PC.

908

NAME

Name sent from PC is too long.

909

RANGE

Numeric Parameter sent from PC is out of range.

920

RS232

Error receiving RS232 data from PC.

921

TIMEOUT

Timed out receiving message from PC.

922

CRC

Message received from PC has incorrect CRC

 

 

value.

923

ILL MSG

Illegal message received from PC.

NOTE: If ETA can be switched on, but fails to respond to the keyboard, the electronic circuitry may be reset by linking pins 5 and 6 in the PC connector, with a length of wire. This will turn the unit off. Turn the unit on again by pressing ‘ON’. If after this reset ETA still fails to operate correctly, use ‘6’ for Miscellaneous and ‘6’ Software Reset to clear memory. Having carried out a hardware reset as described above the message ‘800 MEMORY’ may be displayed on the next power up, and a software reset should be carried out.