3.Installation

3.1Site requirements

Make sure to read and follow all safety instructions in chapter 2.3.2 ‘Site requirements’!

3.1.1 General site requirements

No direct sunlight.

Electrical power supply within distance (length of power cord = approx. 4 meters - do not use extension cords!).

No draft (caused by air conditioning etc.).

Even floor surface.

Practically vibration-free floor.

Handwheel easily accessible.

Room temperature constantly below +22 °C.

Relative humidity of air maximum 60 %.

High ambient temperature and/or high air humidity negatively affect instrument cooling performance!

3.1.2 Electrical connections

Make sure to read and follow all safety instructions in chapter 2.3.3 ‘Electrical connections’!

3.2Unpacking and installation

Unpacking instructions are always located in a transparent protective envelope on the outside of the instrument shipping crate.

Make sure to read and follow all safety instructions provided in chapter 2.3.1 ‘Transport’ and on the unpacking instructions!

3.2.1 Repacking

We recommend to keep the original shipping crate and the unpacking instructions for the Leica CM3050 S.

For repacking, proceed as per unpacking in- structions, in reverse order.

Available models

Basic instrument

-with sectioning motor

-without specimen cooling

Basic instrument

-without sectioning motor

-with specimen cooling

Basic instrument

-with sectioning motor

-with specimen cooling