2.Safety
2.3Safety instructions: handling and operation of the instrument
2.3.1 Transport
•To avoid severe damage to the instrument by running it while the compressor oil is dis- placed from its regular position:
•Do not tilt the instrument,
only transport in an upright position.
•After transport, wait at least 4 hours be-
fore turning the instrument on.
2.3.2 Site requirements
•Do not operate the instrument in rooms with explosion hazard.
•To ensure proper instrument function main- tain a minimum distance of 10 cm between walls and/or furniture and all sides of the in- strument!
2.3.3 Electrical connections
•Do not use extension cords for connecting the instrument to mains.
•During the
•Ensure uniform current supply according to specifications.
•Therefore, arrange for the electrical in- stallations on site to be checked by a trained professional and make sure any necessary upgrades are installed!
•Have the circuit protected by a fuse of its own!
•Do not connect any other consumers to the same circuit.
•Prior to connecting the instrument to mains, make sure the electrical power supply in your laboratory corresponds to the values indicated on the instrument nameplate (located at the rear of the in- strument)!
12 | Instruction Manual V1.2 - 11/2006 |