7.Disinfection, cleaning and maintenance

7.6Exchanging the fluorescent light lamp

Prior to exchanging the lamp, please read and observe the fol- lowing safety instructions:

‘Maintenance’ (Chapter 2.3.13)

Removal

The lamp (3) is located behind the glare shield (4).

Carefully hold lamp with both of your hands. Slightly pull the lamp downward at the left side to remove it from fastening clip (5).

Pull the lamp towards the left to remove it from its socket (6).

Removing the sliding window

Hold sliding window (1) at grip bar (2), lift slightly and pull out towards yourself.

Installing the replacement lamp

• Hold the lamp in correct position (see draw- ing below) and push towards the right until it locks into the socket.

• Push the lamp slightly upwards until it locks into the clip.

• Put the sliding window back in place.

• Reconnect the instrument to mains and switch it back on.