2.Safety

2.3.9 Infectious/radioactive material

Danger!

Use caution when working with potentially infectious specimens:

-->Risk of infection!

When working with potentially infectious / radioactive specimens:

Wear protective clothes (gloves, protec- tive boots, mask, lab coat), in compliance with radiation safety regulations and/or in-house regulations on handling infec- tious / radioactive material.

When working with radioactive specimens:

Comply with applicable radiation safety regulations!

Dispose of radioactive specimen waste ac- cording to applicable regulations.

2.3.10 Disinfection and cleaning

Prior to disinfection, switch the instrument off and unplug it from mains.

For removal of the microtome from the cryo- chamber see chapter 2.3.11 ‘ Removing the microtome’.

For disinfection, wear protective gear: (gloves, mask, lab coat etc.)!

For disinfection, only use alcohol-based dis- infectants!

Do not use solvents (xylene, acetone etc.) for cleaning or disinfection!

Do not spray disinfectants into the evapora- tor!

-->Risk of icing!

Explosion hazard when working with alcohol: Make sure the premises are appropriately ventilated!

When using disinfectants and detergents, comply with all safety instructions supplied by the manufacturer of the product!

Dispose of waste liquids from disinfection/ cleaning as well as of sectioning waste ac- cording to applicable regulations on disposal of special category waste!

Disinfected accessories must be thoroughly dry when reinserting them into the chamber!

-->Risk of icing!

Make sure the chamber is completely dry before switching the instrument back on:

-->Explosion hazard through alcohol vapors! -->Risk of icing!