6. Troubleshooting, applications tips
6.2 Potential problems - causes and remedies
Problem | Causes | Remedies |
Frost on chamber walls and microtome
-Cryostat is exposed to draft (open windows, doors, air- conditioning).
-Sliding window has been open for a long period of time at a very low chamber temperature.
-Frost
-Eliminate draft sources or change place of installation of the cryostat.
-Close sliding window.
-Consider wearing a mask.
Sections smear | - | Specimen not cold enough. | - | Select lower temperature. |
| - | Knife and/or | - | Wait until knife and/or |
|
| yet cold enough - sections |
| plate have reached chamber |
|
| melt. |
| temperature. |
|
|
|
|
|
Sections splinter | - | Specimen too cold. | - | Select higher temperature. |
Sections not properly flattened | - | Static electricity / draft. | - | Eliminate cause. |
| - | Specimen not cold enough. | - | Select lower temperature. |
| - | Large surface specimen. | - | Trim specimen parallel; in- |
|
|
|
| crease section thickness. |
| - | - | Readjust | |
| - | - | Align correctly. | |
|
| relative to knife edge. |
|
|
| - | Wrong clearance angle | - | Set correct clearance angle. |
|
| selected. |
|
|
| - | Knife blunt or damaged. | - | Use different part of knife or |
|
|
|
| replace knife. |
|
|
|
|
|
Sections not properly flattened | - | Knife and/or | - | Clean with dry cloth or brush. |
despite correct temperature | - | Edge of |
|
|
and correctly aligned |
| aged. | - | Replace |
plate | - | Blunt knife. |
|
|
|
|
| - | Use different part of knife or |
|
|
|
| replace knife. |
48 | Instruction Manual V1.1 02/2000 |