6. Troubleshooting, applications tips

6.2 Potential problems - causes and remedies

Problem

Causes

Remedies

Frost on chamber walls and microtome

-Cryostat is exposed to draft (open windows, doors, air- conditioning).

-Sliding window has been open for a long period of time at a very low chamber temperature.

-Frost build-up caused by breathing into the cryocham- ber.

-Eliminate draft sources or change place of installation of the cryostat.

-Close sliding window.

-Consider wearing a mask.

Sections smear

-

Specimen not cold enough.

-

Select lower temperature.

 

-

Knife and/or anti-roll plate not

-

Wait until knife and/or anti-roll

 

 

yet cold enough - sections

 

plate have reached chamber

 

 

melt.

 

temperature.

 

 

 

 

 

Sections splinter

-

Specimen too cold.

-

Select higher temperature.

Sections not properly flattened

-

Static electricity / draft.

-

Eliminate cause.

 

-

Specimen not cold enough.

-

Select lower temperature.

 

-

Large surface specimen.

-

Trim specimen parallel; in-

 

 

 

 

crease section thickness.

 

-

Anti-roll plate poorly adjusted.

-

Readjust anti-roll plate.

 

-

Anti-roll plate poorly aligned

-

Align correctly.

 

 

relative to knife edge.

 

 

 

-

Wrong clearance angle

-

Set correct clearance angle.

 

 

selected.

 

 

 

-

Knife blunt or damaged.

-

Use different part of knife or

 

 

 

 

replace knife.

 

 

 

 

 

Sections not properly flattened

-

Knife and/or anti-roll plate dirty.

-

Clean with dry cloth or brush.

despite correct temperature

-

Edge of anti-roll plate dam-

 

 

and correctly aligned anti-roll

 

aged.

-

Replace anti-roll plate.

plate

-

Blunt knife.

 

 

 

 

 

-

Use different part of knife or

 

 

 

 

replace knife.