32 Instruction Manual V1.2 - 11/2006
5. Operation
5.3.5 Status display

Each individual parameter is displayed in a certain area of the display screen.

Digits 1 and 2
indicate sectioning stroke position:
Sectioning stroke
(outside sectioning window)
: Retraction stroke
[ ]: Sectioning stroke
(inside sectioning window)
Digits 3 - 6
indicate coarse feed position:
Rear limit position: HOME
Front limit position: STOP
Indication of ‘malfunction’ or
‘handwheel locked’:
Microtome plug loose,
implausible values etc.: ????
Handwheel is locked: LOCKED
Indication ‘LOCKED’ completely
overlays digits 3 through 16,
i.e. while ‘LOCKED’ is displayed, no
other parameters will be displayed
in digits 3 - 16 of the lower line.
Further indications: see chapter 6.1
‘Error messages indicated in the
display screen’
Upper line
Lower line
Digits 7 - 16
indicate the following parameters:
Refrigeration status:
Digit 15 = cryochamber
Digit 16 = specimen
Meaning of
refrigeration
symbols:
• Section thickness
(set value)
Section counter
(actual value)
Indication of total section thickness
(actual value)
Preset counter
Digits 7 - 16:
Scroll via arrow buttons to display
individual parameters
Upper line:
In the menu (see chapter 5.3.4): indication of selected menu item.
Standard indication: Chamber temperature (instruments with specimen cooling: chamber and specimen temperature)
Instruments with chamber cooling only
Standard indication upper line:
Instruments with chamber and specimen cooling
-Refrigeration OFF
Refrigeration ON
Defrosting active
Sectioning
stroke

Sectioning window [
]
Retraction stroke

1 2 3 4 5 6 7 8 9 10111213141516

1 2 3 4 5 6 7 8 9 10111213141516

CT-30°C OT-35°C



HOME 

CT-30°C



HOME -