![](/images/new-backgrounds/1138736/138736219x1.webp)
Safety Testing
nThe Toco display shows 0.
TOCO int, time, date and paper speed are printed on the paper.
3.Gently apply pressure to the syringe plunger. Check that the value on the display and paper shows this change in pressure.
If the test fails, repeat using another transducer. If it still fails, refer to
Chapter 8, “Troubleshooting Flowcharts.”
Note— The Intrauterine Pressure Values can be between
Safety Testing
This section defines the test and inspection procedures applicable to the Series 50 A and Series 50 IP fetal monitors. Use the tables in the following section to determine what test and inspection results must be reported after an installation or repair has been carried out.
nTest Blocks in Table
Test and Inspection Matrix in Table
Warning
Apply the safety test limits set by the local standards and statutes applicable to the country of installation, such as IEC
The safety tests defined in this chapter are derived from local and international standards but may not be equivalent. They are NOT a substitute for local safety testing.
Caution
Successful completion of the patient safety tests does not ensure the correct and accurate functioning of the equipment.
Safety Test | The test procedures outlined in this section are to be used only for verifying the safe |
Procedures | installation or service of the product. “Safety checks at installation refer to safety aspects |
| directly related to the installation and setup activities and not to intrinsic safety features that |
| have already been checked during final acceptance testing at the factory.” |
| [FDA |
| manual). First Edition December 1996. Chapter 14: Storage, Distribution and Installation. |
| This document may be viewed at http://www.fda.gov]. |
| When running safety tests on the peripheral components of the system, refer to the |
100 | Chapter 9 Preventive Maintenance, Care and Cleaning |