Transducers and Cables

IUP Pressure Transducer (CPJ840J5)

Width:

37mm/1.5in.

Height:

21.7mm/0.9in.

Operating Temp:

15 to 40°C.

Pressure range:

-20 to + 300 mm Hg

Max. overpressure:

10,000 mm Hg

Sensitivity:

5µV/V/mmHg

Resonance frequency:

300 Hz typical (transducer and dome)

Max. electrical excitation:

15 V DC or AC

Bridge resistance:

1000 Ohms (input and output)

Non-linearity and hysteresis:

max. 0.5% of full scale

Zero balance:

max. 0.15 mm Hg/°C

Operating temperature range:

+10 to +50°C

Storage temperature range:

-20 to +70°C

Isolation resistance:

min. 1000 MOhms

Leakage current:

max.1.5mA at 250V, 50 Hz

Weight:

24 grams (without cable)