Philips manual M1351A/ M1353A Fetal Monitor Test and Inspection Matrix

Models: M1353A M1351A

1 189
Download 189 pages 27.59 Kb
Page 113
Image 113

Safety Testing

Test and

Key to Table: P = Pass and F = Fail, X = test result value to be recorded.

 

Inspection

 

 

 

 

 

Matrix

 

 

 

 

 

Table 9-3 M1351A/ M1353A Fetal Monitor - Test and Inspection Matrix

 

 

 

 

 

 

 

 

 

 

 

 

What to Record

Test Block

Test or “Inspection” to Perform

Expected Test

 

on Service

Name

Results

 

Record (Philips

 

 

 

 

 

 

 

Personnel only)

 

 

 

 

 

Visual

Inspect the unit, transducers and cables for any

If Yes, Visual test is

 

V:P or V:F

 

damage.

passed.

 

 

 

 

Are they free of damage?

 

 

 

 

 

 

 

 

 

Power On

Power on the unit.

If Yes, Power On test is

 

PO:P or PO:F

 

Does the self-test complete successfully?

passed.

 

 

 

 

 

 

 

 

Performance

Perform the quick test and parameter test!

If Yes, Performance

 

P:P or P:F

 

Do these tests complete without errors?

Test is passed.

 

 

 

 

 

 

 

Safety(2)

Protective Earth.

With mains cable:

 

S2:P/x or

 

See -104- Safety (2)

Maximum impedance

 

S2:F/x

 

 

= x (<= 200 mOhms)

 

 

 

 

 

 

 

Safety(3)

Enclosure Leakage Current - Normal Condition.

Maximum leakage cur-

 

S2:P/x4 or

 

See -104- Safety (4)

rent = x1

 

S2:F/x4

 

 

(<= 100A)

 

 

 

Enclosure Leakage Current - S.F.C Open Supply.

Maximum leakage cur-

 

S2:P/x2 or

 

See -104- Safety (5)

rent = x2

 

S2:F/x2

 

 

(<= 500A)

 

 

 

 

(Note: maximum leak-

 

 

 

 

age current in the US =

 

 

 

 

300 A)

 

 

 

Enclosure Leakage Current - S.F.C Open Earth.

Maximum leakage cur-

 

S2:P/x3 or

 

See -104- Safety (6)

rent = x3

 

S2:F/x3

 

 

(<= 500A)

 

 

 

 

(Note: maximum leak-

 

 

 

 

age current in the US =

 

 

 

 

300 A)

 

 

 

 

 

 

 

 

Safety S (3)

M1353A only:

 

 

 

S3(S):P/x or

 

ONLY TOCO-input tested, Metron Testconn.

 

 

 

S3(S):F/x

 

Maximum Leakage

 

 

 

required. (Metron order # 19528 and Philips

 

 

 

Adapter M1940A)

current = x

 

 

 

Patient Leakage Current - S.F.C. Mains on

(<=50

A @ 250V or

 

 

 

Applied Part.

A @ 120V)

 

 

 

See -104- Safety (3)

<= 20

 

 

 

 

 

 

 

 

 

 

 

 

 

Chapter 9 Preventive Maintenance, Care and Cleaning

103

Page 113
Image 113
Philips manual M1351A/ M1353A Fetal Monitor Test and Inspection Matrix

M1353A, M1351A specifications

Philips M1351A and M1353A are advanced medical monitoring devices designed to enhance patient care in various healthcare settings. These monitors are widely recognized for their precision, reliability, and user-friendly interface, making them essential tools for clinicians and healthcare providers.

The Philips M1351A is a compact portable monitor specifically tailored for adult, pediatric, and neonatal care. It is equipped with multiple physiological parameters, enabling healthcare professionals to track vital signs such as heart rate, respiratory rate, non-invasive blood pressure, and temperature. One of its standout features is the ability to display real-time waveforms, providing clinicians with instant access to critical patient information.

On the other hand, the Philips M1353A builds on the M1351A's robust capabilities, integrating advanced features that enhance monitoring capabilities. This model includes additional measurement modalities such as invasive blood pressure and capnography, allowing for comprehensive patient management. Its larger screen provides clear, easy-to-read data, and its intuitive interface ensures quick navigation through the various monitoring functions.

Both monitors utilize state-of-the-art technologies to ensure accurate and reliable measurements. The smart algorithm used in these devices minimizes the possibility of false readings, providing healthcare professionals with the confidence needed for effective clinical decision-making. The Philips M1351A and M1353A also feature an integrated alarm system that notifies caregivers of critical changes in patient conditions, ensuring timely interventions.

In terms of connectivity, these models facilitate seamless integration with hospital networks and electronic medical records (EMR) systems. This interoperability enhances the overall workflow in a clinical environment, allowing patient information to be shared efficiently amongst healthcare teams.

Additionally, the Philips M1351A and M1353A are designed with patient comfort in mind. They are lightweight and portable, making them suitable for diverse environments, from emergency departments to patient transport. Their rugged design ensures durability in high-paced settings, meeting the rigorous demands of modern healthcare.

In summary, the Philips M1351A and M1353A monitors serve as essential tools in contemporary medical practice, offering comprehensive monitoring capabilities, advanced technologies, and a user-centric design. These features collectively contribute to improved patient outcomes and enhanced overall healthcare delivery.