Philips M1353A, M1351A manual Recorder

Models: M1353A M1351A

1 189
Download 189 pages 27.59 Kb
Page 44
Image 44

Configuring the Monitor Using a PC

Adjusting the The Recorder Heat-Adjust is to compensate for different tolerances in the paper.

Recorder

Series 50 Fetal Monitor Configuration and Service Software Rev A.nn.nn

 

Recorder Adjustments:

 

Recorder Heat Adjust:

Actual Value is:

 

Selected Value is

 

 

 

……………

 

Min

Max

Recorder Zero Adjust:

Actual Value is:

 

Selected Value is

 

 

 

……………

 

Left

Right

Return to Main Menu

 

 

MESSAGE:

Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!

To change a setting:

1.Using the appropriate keys, move the cursor to the selection you require.

2.Press Enter.

To return to the Main Menu:

1.Move the cursor to Return to Main Menu.

2.Press Enter.

Example

On the screen above, the actual value of the Recorder Zero-Adjustis closer to the left side. If this is not printed exactly on 20 on the Toco channel, then you must realign the Selected Value is: to the right or to the left and then press Enter. The program will adjust the recorder accordingly.

Note— The actual values on your monitor may be different from those shown.

34

Chapter 4 Configuring the Monitor

Page 44
Image 44
Philips M1353A, M1351A manual Recorder