If the batteries are not replaced when necessary, the time and date return to their default values: when you switch on the Monitor the time is set to 0:00 and date to 04:04:44. Replace batteries with two alkaline AA size 1.5 Volt batteries (recommended type Duracell AA 1.5 Volt Part Number 1420-0393CP). Follow these instructions to replace the batteries:

1.Switch off the Monitor and disconnect it from the main power supply.

2.Press the handle of the paper table and open it by pulling the table forward - lift out any paper.

3.Open the battery cover.

4.Replace the batteries (noting their polarity).

5.Close the battery cover, load paper into the tray and push the paper table closed.

6.Reset the time and date (see Chapter 5, “Setting Time, Date, and Paper Speed.”).

Chapter 11 Replacing Parts