B

Safety and Environment

Safety Information

Follow these guidelines for maximum patient safety:

Protective Earth Check each time before use that the Monitor is in perfect working order and properly grounded.

Warning

Shock hazard: any interruption of the protective earth grounding will cause a potential shock hazard that could result in serious personal injury. Whenever it is ikely that the protection has been impaired, the Monitor must be made inoperative andlbe secured against any unintended operation.

l

l

To protect hospital personnel and the patient, the cabinet must be grounded. Accordingly, the Monitor is equipped with a 3-wire power cable that grounds it to the power line ground when plugged into an appropriate 3-wire receptacle. Do not use a 3-wire to 2-wire adapter with the Monitor.

Position the patient cable so that it does not come into contact with any other electrical equipment. The cable connecting the patient to the Monitor must be free of electrolyte.

Make sure that the Monitor is free from condensation during operation. This can form when equipment is moved from one building to another, and is exposed to moisture and differences in temperature.

Warning

There is a possible explosion hazard if the Monitor is used in the presence of flammable anesthetics. Always disconnect the Monitor from the power supply before removing the top cover. If servicing requires that the power be on while the cover is removed, proceed with extreme caution. Failure to do so can result in serious injury.

Patient Safety

The Monitor should only be used by, or under the direct supervision of, a licensed physician or other health care practitioner who is trained in the use of fetal heart rate monitors and in the interpretation of fetal heart rate traces.

Appendix B Safety and Environment

171

Page 181
Image 181
Philips M1351A, M1353A manual Safety Information, Patient Safety

M1353A, M1351A specifications

Philips M1351A and M1353A are advanced medical monitoring devices designed to enhance patient care in various healthcare settings. These monitors are widely recognized for their precision, reliability, and user-friendly interface, making them essential tools for clinicians and healthcare providers.

The Philips M1351A is a compact portable monitor specifically tailored for adult, pediatric, and neonatal care. It is equipped with multiple physiological parameters, enabling healthcare professionals to track vital signs such as heart rate, respiratory rate, non-invasive blood pressure, and temperature. One of its standout features is the ability to display real-time waveforms, providing clinicians with instant access to critical patient information.

On the other hand, the Philips M1353A builds on the M1351A's robust capabilities, integrating advanced features that enhance monitoring capabilities. This model includes additional measurement modalities such as invasive blood pressure and capnography, allowing for comprehensive patient management. Its larger screen provides clear, easy-to-read data, and its intuitive interface ensures quick navigation through the various monitoring functions.

Both monitors utilize state-of-the-art technologies to ensure accurate and reliable measurements. The smart algorithm used in these devices minimizes the possibility of false readings, providing healthcare professionals with the confidence needed for effective clinical decision-making. The Philips M1351A and M1353A also feature an integrated alarm system that notifies caregivers of critical changes in patient conditions, ensuring timely interventions.

In terms of connectivity, these models facilitate seamless integration with hospital networks and electronic medical records (EMR) systems. This interoperability enhances the overall workflow in a clinical environment, allowing patient information to be shared efficiently amongst healthcare teams.

Additionally, the Philips M1351A and M1353A are designed with patient comfort in mind. They are lightweight and portable, making them suitable for diverse environments, from emergency departments to patient transport. Their rugged design ensures durability in high-paced settings, meeting the rigorous demands of modern healthcare.

In summary, the Philips M1351A and M1353A monitors serve as essential tools in contemporary medical practice, offering comprehensive monitoring capabilities, advanced technologies, and a user-centric design. These features collectively contribute to improved patient outcomes and enhanced overall healthcare delivery.