Transducers and Cables

DECG Transducer (M1357A)

MECG Transducer (M1359A)

DECG/MECG Patient Module (M1364A)

Input Impedance:

>10MΩ (differential, dc to 50/60Hz).

CMRR:

>110dB (with patient cable, 51.5kΩ

/0.047µF imbalance at line

 

frequency).

 

Noise:

<4µVp (referred to input with 25kΩ

).

Contact Potential

 

 

Tolerance:

±400mV.

 

Input Voltage Range:

20µVp to 3mVp.

 

Patient Leakage Current: <10µArms @ 120V/60Hz.

 

Patient Auxiliary Current:<0.1µA (dc).

 

Dielectric Strength:

1500Vrms (spark-gap protected).

 

Weight:

185g/6.5oz.

 

Cable Length:

2.5m/8ft 2in or 0.7m/2ft 3in.

 

Input Impedance:

>10MΩ (differential, dc to 50/60Hz).

CMRR:

>90dB (with patient cable, 51.5kΩ

/0.047µF imbalance at line

 

frequency).

 

Noise:

<4µVp (referred to input with 25kΩ

).

Contact Potential

 

 

Tolerance:

±400mV.

 

Input Voltage Range:

80µVp to 4mVp.

 

Patient Leakage Current: <10µArms @ 120V/60Hz.

 

Patient Auxiliary Current:<0.1µA (dc).

 

Dielectric Strength:

1500Vrms (spark-gap protected).

 

Weight:

175g/6.2oz.

 

Cable Length:

2.5m/8ft 2in.

 

The patient module has a 7-pin ECG connector into which you can plug either DECG cable (M1362A or B) or MECG cable (M1363A).

Overall length: 2706mm (+30, -100mm)

Length of free cable: 2618mm (+30, -100mm)

Weight: 120 grams

Size: 88x42x30mm

Socket: DECG or MECG connection