Philips M1351A, M1353A Service Philosophy, Performance Assurance Tests, Overview, Service Tests

Models: M1353A M1351A

1 189
Download 189 pages 27.59 Kb
Page 75
Image 75

7 Tests and Error Messages

Service Philosophy

Any errors produced are marked on the trace paper with the symbol and a specific error message is shown on the monitor display. In addition, the errors are reported on the system error log (see page 38 for instructions on how to print the error log). Read the error messages and system error log to determine whether the fault lies in the hardware or firmware. These problems can be repaired by board replacement or unit exchange. Do not attempt component-level board repair.

Overview of the

The system contains resident tests to enable monitoring of system status and errors that

Service Tests

occur.

n

n

n

Power-On Self Test (see page 65)

This is invoked whenever the system is switched on.

Performance Assurance Tests

Parameter Test (see page 68)

This checks the processing of the signal from the transducers.

Quick Test (see page 66)

This checks the display, recorder and complete instrument hardware and software.

Service Tests

These are invoked by connecting the monitor to a PC and using a software service tool to assess, test and configure the system.

Cyclic Test (see page 35)

Permanent Test (see page 70)

Safety Tests

These test the electrical safety of the monitoring system and must be carried out after upgrades and repairs. You must use a safety testing tool.

Instrument Safety Test (see page 105)

System Safety Test (see page 105)

Safety Test Blocks (see page 100)

Performance Assurance Tests

Self Test

The monitor automatically performs a basic-level self test when you switch it on. There are

 

two possible types of error that you might see. A fatal error prevents the monitor from

 

functioning. A non-fatal error allows you to continue to work but warns you of a problem

 

thatlmust be resolved swiftly.

 

Ifna non-fatal error occurs (for example, if the batteries are low):

 

An error message is displayed for approximately 13 seconds.

Chapter 7 Tests and Error Messages

65

Page 75
Image 75
Philips M1351A, M1353A manual Service Philosophy, Performance Assurance Tests, Overview, Service Tests, Self Test

M1353A, M1351A specifications

Philips M1351A and M1353A are advanced medical monitoring devices designed to enhance patient care in various healthcare settings. These monitors are widely recognized for their precision, reliability, and user-friendly interface, making them essential tools for clinicians and healthcare providers.

The Philips M1351A is a compact portable monitor specifically tailored for adult, pediatric, and neonatal care. It is equipped with multiple physiological parameters, enabling healthcare professionals to track vital signs such as heart rate, respiratory rate, non-invasive blood pressure, and temperature. One of its standout features is the ability to display real-time waveforms, providing clinicians with instant access to critical patient information.

On the other hand, the Philips M1353A builds on the M1351A's robust capabilities, integrating advanced features that enhance monitoring capabilities. This model includes additional measurement modalities such as invasive blood pressure and capnography, allowing for comprehensive patient management. Its larger screen provides clear, easy-to-read data, and its intuitive interface ensures quick navigation through the various monitoring functions.

Both monitors utilize state-of-the-art technologies to ensure accurate and reliable measurements. The smart algorithm used in these devices minimizes the possibility of false readings, providing healthcare professionals with the confidence needed for effective clinical decision-making. The Philips M1351A and M1353A also feature an integrated alarm system that notifies caregivers of critical changes in patient conditions, ensuring timely interventions.

In terms of connectivity, these models facilitate seamless integration with hospital networks and electronic medical records (EMR) systems. This interoperability enhances the overall workflow in a clinical environment, allowing patient information to be shared efficiently amongst healthcare teams.

Additionally, the Philips M1351A and M1353A are designed with patient comfort in mind. They are lightweight and portable, making them suitable for diverse environments, from emergency departments to patient transport. Their rugged design ensures durability in high-paced settings, meeting the rigorous demands of modern healthcare.

In summary, the Philips M1351A and M1353A monitors serve as essential tools in contemporary medical practice, offering comprehensive monitoring capabilities, advanced technologies, and a user-centric design. These features collectively contribute to improved patient outcomes and enhanced overall healthcare delivery.