Philips M1351A, M1353A manual 12Refitting the Recorder Board

Models: M1353A M1351A

1 189
Download 189 pages 27.59 Kb
Page 155
Image 155

Recorder Board

Recorder Board

To remove the recorder board:

Remove the top cover.

1.Disconnect the cables G, H, I, K, and L.

2.Remove the three screws (1 x S1, 2 x S2). The S2 screws are longer than the S1 screw.

3.Remove the cable J.

4.Lift out the board.

To refit the recorder board:

1.Slide the board into the slot. To give a good electrical contact, the board is tensioned when screwed down.

2.Refit the screws and screw down the board.

Figure 11-12Refitting the Recorder Board

After refitting the board, carry out the Parameter Test (see page 68).

Chapter 11 Replacing Parts

145

Page 155
Image 155
Philips M1351A, M1353A manual 12Refitting the Recorder Board