Philips M1351A, M1353A manual Accessory Testing, Testing Toco Transducers

Models: M1353A M1351A

1 189
Download 189 pages 27.59 Kb
Page 107
Image 107

Accessory Testing

Accessory Testing

Testing Toco

Transducers

Visual check Ensure that there are no cracks in the transducer dome, cable or connector plug.

Electrical check 1. Switch on the monitor and the recorder.

2.Check configuration setting C11 (External TOCO gain) is set to 100% (C11 = 0): See “Configuring the Monitor” on page 32.

3.Connect the transducer to the Toco socket. When you connect the transducer: nThe Toco display shows 20.

TOCO ext, time, date and paper speed are printed on the paper.

4.Lay the transducer face up on a flat surface for a few seconds.

5.Press the Toco Baseline key to re-adjust the Toco display to 20.

6.Turn the transducer over so that the button is face down on the flat surface. Hold the cable at a point 25 cm from the transducer and ensure that the transducer touches the flat surface only with the button and that the transducer is parallel to the flat surface.

The Toco display should read between 40 to 50 units.

paitcoc30

Figure 9-1 Testing a Toco Transducer

If the test fails, repeat using another transducer. If it still fails, refer to

Chapter 8, “Troubleshooting Flowcharts.” After the test, you must zero the system by pressing the Toco Baseline Key.

Note— The external Toco recorder can be between 0 and 127 units.

Chapter 9 Preventive Maintenance, Care and Cleaning

97

Page 107
Image 107
Philips M1351A, M1353A manual Accessory Testing, Testing Toco Transducers

M1353A, M1351A specifications

Philips M1351A and M1353A are advanced medical monitoring devices designed to enhance patient care in various healthcare settings. These monitors are widely recognized for their precision, reliability, and user-friendly interface, making them essential tools for clinicians and healthcare providers.

The Philips M1351A is a compact portable monitor specifically tailored for adult, pediatric, and neonatal care. It is equipped with multiple physiological parameters, enabling healthcare professionals to track vital signs such as heart rate, respiratory rate, non-invasive blood pressure, and temperature. One of its standout features is the ability to display real-time waveforms, providing clinicians with instant access to critical patient information.

On the other hand, the Philips M1353A builds on the M1351A's robust capabilities, integrating advanced features that enhance monitoring capabilities. This model includes additional measurement modalities such as invasive blood pressure and capnography, allowing for comprehensive patient management. Its larger screen provides clear, easy-to-read data, and its intuitive interface ensures quick navigation through the various monitoring functions.

Both monitors utilize state-of-the-art technologies to ensure accurate and reliable measurements. The smart algorithm used in these devices minimizes the possibility of false readings, providing healthcare professionals with the confidence needed for effective clinical decision-making. The Philips M1351A and M1353A also feature an integrated alarm system that notifies caregivers of critical changes in patient conditions, ensuring timely interventions.

In terms of connectivity, these models facilitate seamless integration with hospital networks and electronic medical records (EMR) systems. This interoperability enhances the overall workflow in a clinical environment, allowing patient information to be shared efficiently amongst healthcare teams.

Additionally, the Philips M1351A and M1353A are designed with patient comfort in mind. They are lightweight and portable, making them suitable for diverse environments, from emergency departments to patient transport. Their rugged design ensures durability in high-paced settings, meeting the rigorous demands of modern healthcare.

In summary, the Philips M1351A and M1353A monitors serve as essential tools in contemporary medical practice, offering comprehensive monitoring capabilities, advanced technologies, and a user-centric design. These features collectively contribute to improved patient outcomes and enhanced overall healthcare delivery.