Configuring the Monitor Using Pushbuttons

Table 4-3 NST Timer Paper-Out-Alert

Options

NST Timer

NST automatic

Paper-Out-Alert

recorder off

 

 

 

 

 

 

 

0

 

 

 

 

1

 

 

 

 

2

 

 

 

 

3

 

 

 

 

4

 

 

 

 

5

 

 

 

 

Note— After you put paper into the recorder, let it run for a few pages (use the advance key) to allow the paper to align itself correctly to the right side.

Examples: How to Change the Time Format and IUP Format using Pushbuttons

Example 1:

To change the time format from AM/PM format to 24-hour format, do the following:

1.Disconnect all transducers from the monitor and disconnect or switch off Telemetry.

2.While pressing , press Test.

3.Press the key to change the setting to 1.

4.Press Test to store the new settings.

Example 2:

To change the IUP format from mmHg to kPa, do the following:

Disconnect all transducers from the monitor and disconnect or switch off Telemetry.

While pressing

, press Test.

Press

repeatedly until C03 is displayed.

Press the

 

key to change the setting to 1.

Press Test to store the new settings.