Tests and Error Messages

Operator Error Messages

Table 7-2 Operator Error Messages

Message

Display

Cause and Solution

 

 

 

 

Err 11

US/US1

 

Wrong transducer in the US1/US socket.

 

 

 

Connect the correct transducer.

 

 

 

Err 11

US2/ECG

Wrong transducer in the US2/ECG socket.

 

 

 

Connect the correct transducer.

Err 21

Toco

 

Wrong Toco transducer in the Toco socket.

 

 

 

Connect the correct transducer.

 

 

 

 

Err 41

US

 

US/MECG Combi transducer not allowed.

 

ECG/US2

Remove the transducer.

 

 

 

 

Err 9

US1/US

 

Invalid Telemetry mode.

 

Toco

 

Check the cable from the Telemetry Receiver and, if necessary, replace it. Check all connections.

 

US2/ECG

 

Err 16

US1/US

 

Wrong pairing of Telemetry and transducers.

 

Toco

 

Either disconnect the transducers or switch off the Telemetry Receiver.

 

US2/ECG

 

 

 

 

 

nop

ECG

 

No contact or poor contact between any of the contacts of the system. Check all connections, espe-

 

 

 

cially the legplate connector electrode.

 

 

 

Use a new spiral electrode.

 

 

 

 

Err PAP2/

 

 

Incorrect type of paper loaded.

Err 18

 

 

Replace with recommended paper.

 

 

 

Err bAt

3/ Err 19

Battery low or empty of charge.

 

 

 

Change the batteries as soon as possible. If you do not change the batteries, your specific settings will return

 

 

 

to their default values when the monitor is switched off (for example, the date is set to 4.4.44).

 

 

 

 

Error 601

 

 

Paper speed.

5

 

 

Check that correct paper is used.

 

 

 

Check the speed by timing how long it takes for the paper to advance 1cm:

 

 

 

60 seconds = 1cm/min

 

 

 

30 seconds = 2cm/min

 

 

 

20 seconds = 3cm/min

 

 

 

Contact your Philips Service Engineer or Response Center if the speed is incorrect.

 

 

 

 

Err PAP 30-240

4

Incorrect type of paper loaded.

 

Load paper with 50-210 scale or change the paper format setting of the monitor.

Err 6025

 

 

 

 

 

Err PAP 50-210

4

Incorrect type of paper loaded.

 

Load paper with 30-240 scale or change the paper format setting of the monitor.

Err 6025

 

 

 

 

 

Err 706

 

 

Modem not responding.

 

 

 

The Modem is either incorrectly installed, faulty or not an approved model.

 

 

 

 

Err 776

 

 

Telephone transmission failure.

 

 

 

Line busy or faulty, wait short time and send information again.

Err 5326

 

 

Modem Interface board failure.

 

 

 

Board failure will prevent data transmission from taking place and may possibly generate Err 70 and

 

 

 

Err 77 messages.

1.See the Instructions for Use for your monitor for the types of transducers that can be connected.

2.Displayed for ten seconds after the wrong paper is detected.

3.Displayed for ten seconds when the monitor is first switched on.

4.Printed every ten minutes.

5.Printed every three pages.

6.Only applicable with the Modem Interface Module fitted.

Chapter 7 Tests and Error Messages

69

Page 79
Image 79
Philips M1351A, M1353A manual Operator Error Messages

M1353A, M1351A specifications

Philips M1351A and M1353A are advanced medical monitoring devices designed to enhance patient care in various healthcare settings. These monitors are widely recognized for their precision, reliability, and user-friendly interface, making them essential tools for clinicians and healthcare providers.

The Philips M1351A is a compact portable monitor specifically tailored for adult, pediatric, and neonatal care. It is equipped with multiple physiological parameters, enabling healthcare professionals to track vital signs such as heart rate, respiratory rate, non-invasive blood pressure, and temperature. One of its standout features is the ability to display real-time waveforms, providing clinicians with instant access to critical patient information.

On the other hand, the Philips M1353A builds on the M1351A's robust capabilities, integrating advanced features that enhance monitoring capabilities. This model includes additional measurement modalities such as invasive blood pressure and capnography, allowing for comprehensive patient management. Its larger screen provides clear, easy-to-read data, and its intuitive interface ensures quick navigation through the various monitoring functions.

Both monitors utilize state-of-the-art technologies to ensure accurate and reliable measurements. The smart algorithm used in these devices minimizes the possibility of false readings, providing healthcare professionals with the confidence needed for effective clinical decision-making. The Philips M1351A and M1353A also feature an integrated alarm system that notifies caregivers of critical changes in patient conditions, ensuring timely interventions.

In terms of connectivity, these models facilitate seamless integration with hospital networks and electronic medical records (EMR) systems. This interoperability enhances the overall workflow in a clinical environment, allowing patient information to be shared efficiently amongst healthcare teams.

Additionally, the Philips M1351A and M1353A are designed with patient comfort in mind. They are lightweight and portable, making them suitable for diverse environments, from emergency departments to patient transport. Their rugged design ensures durability in high-paced settings, meeting the rigorous demands of modern healthcare.

In summary, the Philips M1351A and M1353A monitors serve as essential tools in contemporary medical practice, offering comprehensive monitoring capabilities, advanced technologies, and a user-centric design. These features collectively contribute to improved patient outcomes and enhanced overall healthcare delivery.