Philips M1351A System, Frequency, Repetition Rate, Ultrasound Intensity, Signal Range, Mechanism

Models: M1353A M1351A

1 189
Download 189 pages 27.59 Kb
Page 25
Image 25

Monitor

Ultrasound Mode

System:

Pulsed Doppler oscillation.

 

Frequency:

998.4 kHz.

 

Repetition Rate:

3.2 kHz.

 

Ultrasound Intensity:

1.5mW/cm² average for each of the seven active surfaces.

DECG and MECG See Specifications for Transducers and Cables on page 17.

Mode

 

 

External Labor

Signal Range:

0 to 127 units.

 

Offset Compensation:

±200 units.

Internal Labor

Signal Range:

-99 to +127 mmHg.

 

Patient Leakage Current:

10 µArms.

 

Sensitivity:

40 µV/V/mmHg (M1348A).

 

 

5 µV/V/mmHg (M1334A and CPJ840J5).

Recorder

Mechanism:

3-channel, high-resolution (8 dots/mm) thermal array recorder

 

 

with paper-end detection.

 

Paper Speeds:

1, 2 or 3 cm/min.

 

Recording Time Per Pack

 

 

of Paper:

1 cm/min (25 h).

 

 

2 cm/min (12 h 30 min).

 

 

3 cm/min (8 h 20 min).

 

Paper Advance Speed:

24 cm/min (with automatic stop at the paper-end mark).

 

Annotation:

Time of day, date, and paper speed are printed automatically

 

 

every ten minutes. Monitoring mode is printed with every

 

 

alteration of parameter.

 

Paper:

Fanfold paper with numbered pages.

 

FHR Scale:

USA: 30 to 240 bpm @ 30 bpm/cm.

 

 

Other countries: 50 to 210 bpm @ 20 bpm/cm.

 

Labor Scale:

0 to 100 units @ 25 units/cm.

Chapter 2 Technical Specifications

15

Page 25
Image 25
Philips M1351A System, Frequency, Repetition Rate, Ultrasound Intensity, Signal Range, Offset Compensation, Sensitivity