Recorder Sensing Assembly

and slide it into position.

5.Refit the recorder chassis into the top cover.

6.Refit the RFI bracket.

7.Refit the countersunk screws.

8.Reconnect the printhead connector.

9.Refit the recorder board.

10.Reconnect the recorder-board cables.

Note— Do not fit a screw into the vacant screw position near the stepper motor.

This screw position is reserved.

After replacing the thermal printhead, carry out the System Test (see page 105).

Recorder Sensing Assembly

To remove the recorder sensing assembly:

1.Remove the top cover and Recorder Board.

2.Remove the RFI bracket.

3.Remove the screw securing the recorder sensing assembly.

4.Disconnect the cable.

To refit the recorder sensing assembly, reverse the above procedure.

Caution

The use of recorder paper that is not approved by Philips can damage the

Monitor. This type of damage will not be covered by warranty.