Setting the Hi Vacuum Sensor Delay

Hi vac sensor delay is the time that a high vacuum condition must exist before it is sensed by the controller. This delay eliminates the annoying Hi Vacuum errors which self-clear or are erroneous. This delay defaults to 3.0 seconds and should not be altered unless you are instructed to do so by the Service Department.

HI VAC SENSOR DELAY?

NO/YES

At the above message prompt:

1. Press the YES key to enter the Hi Vacuum Sensor Delay Time menu.

SELECT PUMP 1 OR 2

USE KEYPAD TO ENTER

At the above message prompt:

2.Select the pump for altering vacuum sensor delay. The following screen displays:

HIVAC DELAY: 3.000s

SETTING FOR PUMP n

Using the keypad:

3. Enter the new time and press the ENTER key to save the new setting.

2–Pump 12-Station Controllers

Chapter 8: Service Only Information