Setting the Hi Vacuum Vent Delay

Hi vacuum vent delay is the vent time initiated by a high vacuum condition. This delay defaults to 10.0 seconds and should not be altered unless you are instructed to do so by the Service Department.

HI VAC VENT DELAY?

NO/YES

At the above message prompt:

1. Press the YES key to enter the Hi Vacuum Vent Delay Time menu.

SELECT PUMP 1 OR 2

USE KEYPAD TO ENTER

At the above screen prompt:

2.Select the pump for altering vacuum vent delay. The following screen displays.

HV VENT DLY: 10.000s

SETTING FOR PUMP n

Using the keypad:

3. Enter the new time and press the ENTER key to save the new setting.

2–Pump 12-Station Controllers

Chapter 8: Service Only Information

62