Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

9.3.4.3 Status Output Lines, External Contact Closures

The Status lines consist of 2 active input lines, and 12 active output lines. Additional circuits are present on the board but currently unused. Individual lines are set or cleared under CPU control depending on the assigned alarm condition. The CPU also monitors the 2 input lines for remote calibration commands. The status inputs and outputs are terminated at the rear panel, see Rear Panel Connector Board schematic diagram in the Appendix.

The output lines are opto-coupled NPN transistors which can pass 50 ma max of direct current with a voltage of 30 VDC max.

The input lines are optically coupled with inputs pulled up to +5 VDC. External contacts can be contact closures or open collector transistor contacts. DO NOT apply any voltage, since +5 VDC is supplied internally.

Individual status lines can be set or cleared using the DIAGNOSTIC mode SIGNAL I/O. This can be useful for simulating fault conditions in the analyzer to see if external circuitry is working correctly. See Table 5-9 for pin assignments.

9.3.4.4 4-20 mA Current Output

4-20 mA current loop option replaces the voltage output of the instrument with an isolated 4-20 mA current output. The current outputs come out on the same terminals that were used for voltage outputs, see Figure 2-2. It is programmable for 4-20mA or 0-20mA and has a 1500 V common mode voltage isolation and 240 V RMS normal mode voltage protection.

Vloop = 28V max which is sufficient to drive up to a 1000 ohm load.

9.3.5 Power Supply Module

The Power Supply Module consists of several subassemblies described in Table 9-9.

9-47

Page 183
Image 183
Teledyne 200AU instruction manual Status Output Lines, External Contact Closures, 4.4 4-20 mA Current Output

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.