Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

Table 5-19: RS-232 Calibration Examples

Action

RS-232 Commands

Comments

Zero Calibration

C ZERO

Z/S valves switched to admit

 

C COMPUTE ZERO

zero gas. Zero cal in Single

 

C EXIT

Range mode

 

 

 

Zero Calibration of low range -

C ZERO 1

Z/S valves switched to admit

AutoRange Enabled

C COMPUTE ZERO

zero gas. Zero calibration of

 

C EXIT

low range in Auto Range mode

 

 

 

Span Calibration of high range -

C SPAN 2

Z/S valves switched to admit

AutoRange Enabled

C COMPUTE SPAN

span gas. Span calibration of

 

C EXIT

high range in Auto Range

 

 

mode

 

 

 

Zero Calibration with Dynamic

C ZERO

Z/S valves switched to admit

Calibration enabled

C EXIT

zero gas. Instrument is zero

 

 

calibrated if DYN CAL is

 

 

enabled.

 

 

 

Zero Calibration

C ZERO

Z/S valves switched to admit

 

C EXIT

zero gas. Instrument zero is

 

 

just checked, but not changed.

 

 

 

Execute AutoCal Sequence #2

C ASEQ 2

Execute a predefined AutoCal

 

 

Sequence. Executes sequence

 

 

immediately, ignoring time and

 

 

date parameters.

 

 

 

Span Calibration Check

C SPAN

Z/S valves switched to admit

 

C EXIT

span gas. Instrument span is

 

 

just checked, but not changed.

 

 

 

Whenever the analyzer starts or finishes a ZS calibration, it issues a status report to the RS-232 interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise, they will be discarded. Table 5-20 shows the format of the text of the calibration messages. An example of an actual sequence of calibration status messages is:

C DDD:HH:MM IIII START MULTI-POINT CALIBRATION

C DDD:HH:MM IIII NOX=xxxxx PPB NO=xxxxx PPB NO2=xxxxx PPB C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION

5-35

Page 73
Image 73
Teledyne 200AU instruction manual 19 RS-232 Calibration Examples, Action RS-232 Commands Comments

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.