Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

Table 9-11: Ozone Generator Control Conditions

Condition

Description

Ozone

Generator

 

 

1

Manual override off (SETUP-MORE-DIAG-OZONE GEN).

OFF

 

 

 

2

Manual override on.

ON

 

 

 

3

Ozone flow below low warning limit for 5 minutes (OFLOW_SET

OFF

 

setup variable, typically <40 cc/min).

 

 

 

 

4

Instrument powered on for more than 30 minutes.

ON

 

 

 

5

Instrument powered off for less than 1 hour and ozone generator was

ON

 

on when instrument was powered off.

 

 

 

 

6

Ozone flow above low warning limit for 0.5 minutes and condition 4

ON

 

or 5 is true.

 

 

 

 

NOTE

The ozone generator is independently controlled in the SIGNAL I/O, OPTIC TEST, and ELECTRICAL TEST diagnostics. After exiting these diagnostics, the ozone generator is restored to the state specified by the above conditions.

9.3.6.1 PermaPure Drier

The PermaPure drier is constructed of 2 concentric tubes. The inner tube is a special material that has an affinity for water vapor. The outer annulus is evacuated by the instrument pump. This creates a concentration gradient causing water in ambient air to diffuse into the outer annulus, thus the air in the inner tube becomes progressively drier as it progresses down the tube.

Due to the large number of connections and fittings on the drier, the most common drier fault is leaks. Before proceding with any other procedures check the drier for leaks.

Occasionally the drier gets contaminated. The manufacturer of the drier recommends replacing the drier rather than trying to clean it. If cleaning is chosen, the following options are available:

1.Dirt - Clean any solids from dryer inlet by brushing. Use clean dry air to blow any loose particles from the inlet. Deionized water or dilute(5-10% conc HCl in deionized water) hydrochloric acid can be passed through the dryer. This should be done only with the dryer and HCl at room temperature.

2.Organic liquids and freons - Rinse the inner and outer tubes using 1,1,1 trichlorethane. Follow the solvent with dry air to purge the solvent.

3.Inorganic salts and oxgenated hydrocarbons can be removed by rinsing with 10% nitric acid at 50° C.

9-53

Page 189
Image 189
Teledyne 200AU instruction manual Ozone Generator Control Conditions, PermaPure Drier

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.