Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

8 MAINTENANCE

8.1 Maintenance Schedule

NOTE

The operations outlined in this chapter are to be performed by qualified maintenance personnel only.

Table 8-1: Preventative Maintenance Schedule

Item

Maintenance Interval

Reference Section

TEST Functions

6 - 12 month intervals

Table 9-1

 

 

 

Zero/Span Calibration

Annually or after repairs

Section 7

 

 

 

Zero/Span Checks

Daily

Section 7, Table 6-2

 

 

 

Particulate Filter

Weekly as needed

Figure 8-1

 

 

 

Filter for PermaPure Drier

Replace every 12 months

Figure 9-7

 

 

 

Reaction Cell Window

Clean annually or as necessary

Section 9.3.8, 8.4, Figure 8-3

 

 

 

Ozone Flow

Check every year

Figure 9-9, Section 9.3.7

 

 

 

Sample Flow

Check every year

Figure 9-9, Section 9.3.7

 

 

 

Moly Converter

Check efficiency every 6

Figure 8-4, Section 8.5

 

months

 

 

 

 

Pneumatic Lines

Examine every 12 months, clean

Figure 8-5, Figure 8-6, Figure 8-7

 

if necessary

 

 

 

 

Factory Calibration

Calibrate each year or after

Section 9.1.6

 

repairs

 

 

 

 

Leak Check

Check every year

Section 8.7

 

 

 

O-rings

Replace if seal is broken

Figure 9-9, Figure 9-10

 

 

 

8-1

Page 119
Image 119
Teledyne 200AU instruction manual Preventative Maintenance Schedule, Maintenance Interval Reference Section

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.