Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

6.Ozone concentration too low

A.Wet air in ozone generator - need to flush out clean, dry

B.Ozone generator driver board failure.

C.Damaged transformer/ozone generator cartridge. The ozone generator cartridge is partially made of glass. Extreme physical shock, such as dropping the generator on the floor may cause the glass cartridge to crack and eventually fail.Sample inlet vent line too short - should be at least 15".

7.Sample exhaust not properly vented, creating a backpressure at the sample inlet port of the instrument. See Figure 2-3 for venting recommendations.

9.2.11 Slow Response

1.Contaminated or dirty sample delivery pneumatics

A.Dirty/plugged sample filter or sample lines

B.Dirty reaction cell

2.Sample inlet line too long.

3.Wrong materials in contact with sample - use glass, stainless steel or Teflon

4.Sample vent line located too far from instrument sample inlet. Locate sample inlet vent near analyzer.

5.Insufficient time allowed for purging of lines upstream of analyzer.

6.Leaking NO/NOx valve.

7.Insufficient time allowed for NO or NO2 cal gas source to become stable.

8.Moly converter temperature too low.

9.Miscalibration, see Table 7-1.

9.2.12 Analog Output Doesn't Agree with Display Concentration

The analog output is proportional to the range. Zero volts output corresponds to zero ppb, and 5 volts corresponds to the maximum range setting in ppb. If this is not observed do the following:

1.V/F board DAC's out of calibration. Do DAC calibration and Factory Calibration.

2.Analog outputs electrically loaded down causing voltage to sag. Could be due to input impedance to chart recorder or data logger being too low or improper grounding. The Recorder and DAS outputs do not have separate output drivers, the problem could be the combined load of both could be too high.

9.3Subsystem Troubleshooting and Adjustments

9.3.1 Computer, Display, Keyboard

9-32

Page 168
Image 168
Teledyne 200AU instruction manual Subsystem Troubleshooting and Adjustments, Slow Response, Computer, Display, Keyboard

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.