Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

The function of each of the default Data Channels is described below:

CONC: Samples NOx, NO and NOy concentration data at one minute intervals and stores an average every hour with a time and date stamp. Readings during calibration and calibration hold off are not included in the data. The last 800 hourly averages are stored.

PNUMTC: Collects sample flow and sample pressure data at five minute intervals and stores an average once a day with a time and date stamp. This data is useful for monitoring the condition of the pump and critical flow orifice(sample flow) and the sample filter(clogging indicated by a drop in sample pressure) over time to predict when maintenance will be required. The last 360 daily averages (about 1 year) are stored.

CALDAT: Logs new slope and offset every time a zero or span calibration is performed, also records the sample concentration reading just prior to performing a calibration.

STABIL: Logs the standard deviation of the last 50 minutes of NOx data, with readings taken 2 minutes apart(i.e. 25 readings taken 2 minutes apart.) The collector operates in “sliding window” fashion with the oldest reading being deleted and a new reading added every 2 minutes. A new value of STABIL is added to the data buffer every 2 minutes. A time and date stamp is recorded for every data point logged.

NOTE:

This Data Channel collects data based on an event (a calibration) rather than a timer. This Data Channel will store data from the last 200 calibrations. This does not represent any specific length of time since it is dependent on how often calibrations are performed. As with all Data Channels, a time and date stamp is recorded for every data point logged.

This is the EPA definition of noise used for equivalency testing. The collector provides a continuous readout of the noise. In order for the data to be meaningful, the instrument must sample a constant concentration of gas for at least 50 minutes.

5-18

Page 56
Image 56
Teledyne 200AU instruction manual

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.