Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

Table 6-2: Example of AutoCal Setup

Step

Action

Comment

1.

Press

This button sequence will cause the AUTOCAL menu to be

 

SETUP-ACAL

displayed.

 

 

 

2.

Press PREV-NEXT

Press PREV-NEXT until SEQ 2 is displayed

 

 

 

3.

Press MODE

Select the MODE menu

 

 

 

3.

Press PREV-NEXT

Press PREV NEXT to scroll to SPAN

 

 

 

4.

Press ENTR

ENTR selects the SPAN MODE

 

 

 

5.

Press SET

Select the SET menu to change the sequence attributes

 

 

 

6.

Press PREV-NEXT

Scroll the SET menu to TIMER ENABLE

 

 

 

7.

Press EDIT

Allows changing the TIMER ENABLE attribute, select ON

 

 

 

8.

Press ENTR

ENTR changes TIMER ENABLE to ON

 

 

 

9.

Press PREV-NEXT

Repeat steps 6-9 for each attribute

 

 

 

10.

Press EXIT

Press the EXIT key to return to upper level menus

 

 

 

6.4 4-20 mA, Current Loop Output

The current loop option replaces the voltage output of the instrument with an isolated 4-20 mA current output. The current outputs come out on the same terminals that were used for voltage outputs, see Figure 2-2. See Troubleshooting Section 9.3.3 for setup and calibration.

6.5 NOy Converter

The NOy Converter Option allows placement of the molybdenum converter at the sample inlet point. This configuration is useful for background studies and where it is desirable to immediately convert certain components of the atmosphere that decompose rapidly, or may be lost on the walls of a normal sample induction system. This group of compounds is sometimes referred to as NOy.

The Option consists of a modified M200AU, a M501Y chassis and remote converter. The pneumatic diagram for this system of components is in Figure 8-7. Because of the nature of this option, there is a separate manual “M501Y – Remotely Mounted Converter” (P/N 02808). The manual covers setup, calibration, and operation of the system.

6-4

Page 82
Image 82
Teledyne 200AU 4-20 mA, Current Loop Output, NOy Converter, Example of AutoCal Setup, Step Action Comment, Setup-Acal

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.