Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

PSM Diagnostic Procedures

The Linear Power Supply board can be tested by checking the DCPS - TEST function on the front panel. It should read 2500 mV ± 200 mV. If the value is outside this range, individual output voltages can be tested on connector P3, see Schematic in the Appendix for pinouts.

The Switching Power Supply output can be tested by observing the temperature of the PMT cold block using the PMT TEMP - TEST function. The temperature should be constant –5 ± 2°C. The output voltage can be observed on J10 of the Switch Board. It should be 15 VDC ± 0.5.

The Switch Board can be tested by observing the diagnostic LEDS along the top edge of the board. The following Table 9-10 describes the typical operation of each LED.

Table 9-10: Power Supply Module LED Operation

No.

Function

Description

1.

NO/NOx Valve

Should switch about every 5 sec.

 

 

On(click sound) = NOx mode

 

 

Off(thud sound) = NO mode

 

 

 

2.

Zero/Span Valve

Should switch ON when CALZ or CALS button is

 

 

pressed.

 

 

 

3.

Sample/Cal Valve

Should switch ON when CALS button is pressed.

 

 

 

4.

Pre-reactor Valve

ON when M200AU in Pre-reactor mode. Happens once

 

 

every 6 NO/NOx cycles or about once per minute.

5.

Ozone Generator Power

Ozone generator power is controlled by:

 

 

1.

Molybdenum converter temp

 

 

2.

Time since last power-up

 

 

The ozone generator will be on if:

 

 

1.

Moly temp is < 100 C and its been more than 30 min

 

 

 

since power up.

 

 

2.

Moly temp between 100-250 C and its been more than

 

 

 

10 min since power up.

 

 

3.

Moly temp > 250 C

 

 

 

6.

Sample flow control block

Should cycle ON-OFF every 20 sec to 2 min. On

 

heater

continuously until up to temp.

 

 

 

7.

Converter Heater

Should cycle ON-OFF every 20 sec to 2 min. On

 

 

continuously until up to temp.

 

 

 

8.

Reaction Cell Heater

Should cycle ON-OFF every 20 sec to 2 min. On

 

 

continuously until up to temp.

 

 

 

 

9-51

Page 187
Image 187
Teledyne 200AU instruction manual PSM Diagnostic Procedures, Power Supply Module LED Operation, Function Description

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.