Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F

Table 9-2: Front Panel Warning Messages

Message

Description

SYSTEM RESET

Analyzer was reset/powered on. This warning occurs every

 

time the instrument is powered up, as in a power failure. It can

 

also occur if the RAM or EEPROM is reset.

 

 

RAM INITIALIZED

RAM was erased. The RAM contains the DAS averages which

 

get erased when the RAM is initialized. It also contains

 

temporary data used by the M200AU to calculate

 

concentrations. No setup variables are stored in the RAM.

 

 

SAMPLE FLOW WARN

The calculated sample flow is outside the hi/low limits. Since

 

the flow is calculated, it probably means the pressure has

 

gotten too low. Flow warnings are caused by a plugged sample

 

inlet or ozone line.

 

 

OZONE FLOW WARNING

Ozone flow out of spec. Warnings occur most often because of

 

loss of vacuum, which causes the ozone flow to go to zero.

 

They also can occur due to a flow sensor failure.

 

 

RxCELL PRESS WARNING

Vacuum out of spec. Warnings are caused by leaks, pump

 

failure or disconnected pump.

 

 

BOX TEMP WARNING

Box temp. out of spec. Instrument fan failure, enclosure

 

temperature failure. Operation of the M200AU in a too warm

 

or cold environment will cause degradation of data quality and

 

shorten the life of the instrument.

 

 

RCELL TEMP WARNING

Reaction cell temp. out of spec. The warning message is most

 

often present during initial warm-up or if the connector to the

 

heaters is not plugged in after dis-assembly. It has also

 

occurred if the thermistor is not in position in the reaction cell.

 

 

PMT TEMP WARNING

PMT temp. out of spec. The PMT temp has its own

 

proportional controller on the preamp board. Warnings might

 

occur during warmup. A warning can occur if the 7 pin

 

connector to the interior of the sensor is not plugged in. The

 

power connector to the PSM should be checked for proper

 

voltage (+15 VDC ± 0.5)

 

 

MOLY TEMP WARNING

Molycon temp. out of spec. The Moly temp is controlled by

 

the CPU. It has a thermocouple with amplifier on the

 

Status/Temp board. Because of the high temperature of the

 

Moly (315° C), the moly temp warning will tend to be the last

 

warning to clear as the instrument is powered on.

 

 

(table continued)

9-9

Page 145
Image 145
Teledyne 200AU instruction manual Front Panel Warning Messages, Message Description

200AU specifications

The Teledyne 200AU is a cutting-edge analytical instrument designed primarily for trace and ultra-trace analysis of elements in various liquid samples. Renowned for its reliability and precision, this inductively coupled plasma mass spectrometer (ICP-MS) is widely utilized in environmental monitoring, food safety, clinical diagnostics, and geochemical research.

One of the distinguishing features of the Teledyne 200AU is its exceptional sensitivity, allowing for the detection of elements at concentrations as low as parts per trillion (ppt). This sensitivity is crucial for applications where even minute traces of contaminants can have significant implications. The instrument also boasts a wide dynamic range, accommodating high-concentration samples without compromising accuracy.

The Teledyne 200AU incorporates advanced technologies that enhance its analytical capabilities. At the core of its design is a robust ICP source, which ensures optimal ionization efficiency. The instrument also features a high-resolution quadrupole mass analyzer, which enables precise separation and identification of ions based on their mass-to-charge ratio. This feature is essential for distinguishing between isotopes and analyzing complex mixtures.

Another notable characteristic of the Teledyne 200AU is its user-friendly interface, which simplifies the operation and data management processes. The software is intuitive, offering sophisticated data processing algorithms that streamline analysis. Users can easily access real-time data visualization, facilitating timely decision-making based on analytical results.

The instrument is built with durability in mind, featuring a compact design that allows for easy integration into various laboratory environments. The Teledyne 200AU is also equipped with state-of-the-art safety features, including enhanced cooling systems and automated shutdown protocols to protect both the instrument and the user.

Moreover, the Teledyne 200AU supports a variety of sample introduction methods, including nebulization and laser ablation, providing flexibility to accommodate different types of samples and experimental needs. Its multi-element capability further enhances its versatility, allowing simultaneous detection of multiple elements in a single analysis.

In summary, the Teledyne 200AU stands out as a premier instrument in the field of mass spectrometry. Its combination of high sensitivity, advanced technology, user-friendly features, and robust design makes it an invaluable tool for scientists and researchers seeking precise elemental analysis across various disciplines.