5.2Adjustment of Height
Note:
Check if the image adjustment for the equipment is performed properly before this adjustment of the RADF. (See the Service Handbook of the applicable equipment.)
[A]Checking
(1)Close the RADF.
(2)Light the exposure lamp.
•Turn the power ON while pressing [0] and [3] simultaneously.
•Key in [267] and then press the [START] button. The exposure lamp is turned ON for a given length of time.
(3)Visually check the gap between platen guide holder "A" and upper surface of the original glass "B" from the left hand side of the equipment. If the value is not within the tolerance, perform the adjustment according to the following procedure.
[Tolerance of the gap] Rear side: 0 - 0.5 mm Front side: 0 mm
0~0.5mm
A
B |
A
B |
0mm