Trane RAUC-C40 manual Standing Vacuum Test, System Pre-StartProcedures, Typical Vacuum Pump Hookup

Models: RAUC-C20 RAUC-C40 RAUC-C25 RAUC-C30 RAUC-C60 RAUC-C50

1 112
Download 112 pages 31.47 Kb
Page 70
Image 70
Standing Vacuum Test

System Pre-Start Procedures

Note: It is unlawful to release refrigerant into the atmosphere. When service procedures require working with refrigerants, the service technician must comply with all Federal, State, and local laws. Refer to the General Service Bulletin MSCU-SB-1 (latest edition).

Standing Vacuum Test

Once 300 microns or less is obtained, close Valve A and leave valves B and C open. This will allow the vacuum gauge to read the actual system pressure. Let the system equalize for approximately 15 minutes. This is referred to as a “standing vacuum test” where, time versus pressure rise. The maximum allowable rise over a 15 minute period is 200 microns. If the pressure rise is greater than 200 microns but levels off to a constant value, excessive moisture is present. If the pressure steadily continues to rise, a leak is indicated. Figure 35 illustrates three possible results of the “standing vacuum test”. If a leak is encounter, repair the system and repeat the evacuation process until the recommended vacuum is obtained. Once the system has been evacuated, break the vacuum with refrigerant, and complete the remaining “Pre-Start Procedures” before starting the unit.

Figure 34. Typical Vacuum Pump Hookup

70

SS-SVX09A-EN

Page 70
Image 70
Trane RAUC-C40, RAUC-C50, RAUC-C30, RAUC-C60 Standing Vacuum Test, System Pre-StartProcedures, Typical Vacuum Pump Hookup