Tripp Lite SMART2500XLHG owner manual Electrostatic Discharge ESD

Page 11

Guidance and Manufacturer’s Declaration cont.

Guidance and Manufacturer’s Declaration—Electromagnetic Immunity

This Medical-Grade UPS is intended for use in the electromagnetic environment specified below. The customer or the user of this Medical-Grade UPS should assure that it is used in such an environment.

 

 

Compliance

Electromagnetic

Immunity Test

IEC 60601 Test Level

Level

Environment - Guidance

Electrostatic Discharge (ESD)

+ / - 6kV contact

Complies

Floors should be wood,

IEC 61000-4-2

+ / - 8kV air

 

concrete or ceramic tile. If

 

 

 

colors are covered with

 

 

 

synthetic material, the

 

 

 

relative humidity should be

 

 

 

at least 30%.

 

 

 

 

Electrical Fast Transient/Burst

+ / - 2kV for power supply

Complies

Mains power quality should

IEC 61000-4-3

lines

 

be that of a typical

 

+ / - 1kV for input/output

 

commercial or hospital

 

lines

 

environment.

Surge

+ / - 1kV differential mode

Complies

Mains power quality should

IEC 61000-4-5

+ / - 2kV common mode

 

be that of a typical

 

 

 

commercial or hospital

 

 

 

environment.

 

 

 

 

Voltage dips, short

<5% UT

Complies

Mains power quality should

interruptions and voltage

(>95% dip in UT) for .5 cycle

 

be that of a typical

variations on power supply

40% UT

 

commercial or hospital

input lines

(60% dip in UT) for 5 cycles

 

environment.

IEC 61000-4-11

70% UT

 

 

 

(30% dip in UT) for 25 cycles

 

 

 

<5% UT

 

 

 

(>95% dip in UT) for 5 seconds

 

 

Power Frequency (50/60 Hz)

3 A/m

Complies

Power frequency magnetic

Magnetic Field

 

 

fields should be at levels

IEC 61000-4-B

 

 

characteristic of a typical

 

 

 

location in a typical

 

 

 

commercial or hospital

 

 

 

environment.

 

 

 

 

Note: UT is the AC mains voltage prior to application of the test level.

Overcurrent Protection

The unit is equipped with 2 20A, branch-rated, circuit breakers.

Battery Pack

The unit comes with a 48VDC, rechargeable sealed lead-acid battery pack rated at 9 amp hours. Your battery should last for several years. If needed, replacement battery packs for this UPS are available through Tripp Lite, model number RBC48V-HGTWR.

11

12-10-307-93-3181.indb 11

11/2/2012 9:27:49 AM

Image 11
Contents Intelligent, Line-Interactive Medical-Grade UPS System UPS Connection Warnings Important Safety InstructionsStatement of Intended Use UPS Location WarningsPackage Contents Turn the UPS on Quick InstallationPlug the UPS into an outlet on a dedicated circuit Plug your equipment Into the UPSUSB and RS-232 Serial Communications Optional InstallationUser Interface EPO Port Connection External Battery ConnectionButtons Front Panel Basic OperationIndicator Lights Front Panel Other UPS Features Rear Panel Guidance and Manufacturer’s Declaration Electrostatic Discharge ESD Storage Storage and ServiceBattery Replacement ServiceConnect new batteries Warranty RegistrationRemove battery cover Remove old batteriesSistema UPS Inteligente, Interactivo de Grado Médico SMART2500XLHGAdvertencias sobre la conexión del UPS Instrucciones de seguridad importantesDeclaración de Uso Indicado Advertencias sobre la ubicación del UPSDesecho del UPS Contenido del EmpaqueAdvertencias sobre la conexión de equipos Advertencias sobre la bateríaEncienda el UPS Instalación rápidaEnchufe el UPS en un tomacorriente en un circuito dedicado Conecte sus equipos En el UPSComunicaciones USB y serie RS-232 Interfaz de UsuarioInstalación opcional Conexión de la Batería Externa Instalación opcional continúaConexión de puerto Botones Panel frontal Operación básicaLuces indicadoras Panel frontal Operación básica continúaOtras funciones del UPS Panel posterior Orientación y Declaración del Fabricante Orientación y Declaración del Fabricante continúa Servicio Almacenamiento y servicioAlmacenamiento Reemplazo de batería Manuel du propriétaire Mises en garde Connexions de lUPS Mises en garde Emplacement de lUPSDirectives de sécurité importantes Énoncé sur lutilisation prévueÉlimination du système d’alimentation sans coupure UPS Contenu de lemballageMises en garde Connexion déquipement Mises en garde BatterieMettre lUPS sous tension Installation rapideBrancher l’UPS dans une sortie de circuit à usage exclusif Connecter votre Équipement à lUPSPorts de communication de série USB et RS-232 Installation en optionInterface utilisateur Raccord du bloc-piles Externe2b Installation en option suiteConnexion au port EPO Boutons Panneau avant Fonctionnement de baseVoyants indicateurs Panneau avant Fonctionnement de base suiteAutres caractéristiques de lUPS panneau arrière Directives et déclaration du fabricant Directives et déclaration du fabricant suite Entreposage Entreposage et serviceRetirer les piles usagées Remplacement de batterieÉloigner, avec précaution, le panneau avant de l’unité UPS Retirer le couvercle des pilesIndb 11/2/2012 92822 AM