SECTION 7—RECLINER
Replacing Threaded Sector
NOTE: For this procedure, refer to FIGURE 7.5.
1.Disconnect the lower cable assembly. Refer to Disconnecting/Connecting Lower Cable Assembly on page 61.
2.Remove the release handle. Refer to Replacing Release Handle on page 58.
3.Remove the existing threaded sector by performing the following:
A.Thread the barrel end of the cable through the larger hole on the threaded sector.
B.Remove existing threaded sector from housing.
4.Install the new threaded sector:
A.Thread barrel end of cable assembly through the larger hole in the new threaded sector.
B.Position the barrel end of cable in smaller hole of new threaded sector.
5.Reinstall the release handle. Refer to Replacing Release Handle on page 58.
6.Squeeze the release handle and verify that the cable is activating the threaded sector.
7.If release handle does not activate threaded sector, perform the following:
A.Make sure barrel end of cable is positioned in smaller hole of new threaded sector.
B.Ensure release handle is installed properly, repeat STEP 7.
C.If release handle still does not activate threaded sector, contact Invacare technical support at the number listed on the back cover.
8.Connect the lower cable assembly. Refer to Disconnecting/Connecting Lower Cable Assembly on page 61.
Larger Hole | Barrel End of Cable | |
(STEPS 3,4) | (STEPS 3,4) | |
Threads |
| |
Support Housing | Threaded Sector | |
(STEPS 3,4) | ||
(STEP 3) | ||
| ||
| Smaller Hole | |
| (STEP 4) |
Cable Assembly
FIGURE 7.5 Replacing Threaded Sector
9000 Series | 64 | Part No. 1076155 |