If the unit does not receive an external power source or the battery does not get recharged immediately, the low battery indicator light will remain on and the performance of the unit will drop off rapidly. Switch to another power source immediately to avoid an interrupted suction procedure.
Adjust the vacuum level from 80 to 550 mm Hg by turning the vacuum regulator knob located on the side of the unit (clockwise to increase vacuum and
NOTE– Gauge is for reference only. If the unit sustains a severe drop, accuracy of the gauge must be checked.
Suction at the patient is automatically obstructed when liquid level reaches the float
NOTE– Always transport unit with vacuum regulator knob rotated fully clockwise in case unit is dropped.
CAUTION– When automatic float
CAUTION– Should fluid be aspirated back into the unit, equipment provider servicing is necessary as possible vacuum pump damage may result.
BATTERY CHARGING
On 7305P series, the units are equipped with a
Once you've determined that your unit is equipped with an internal rechargeable battery, connect the unit to the proper AC or DC power source using the AC adapter or DC power cord. The green external power light shall be illuminated. The yellow charge indicator will be illuminated while the bat- tery is charging. Verify that this illuminates when charging begins. As the charge nears a full charge, the yellow LED may flash on and off for several minutes. This is normal.
NOTE– A discharged battery will require
NOTE– Do not connect the AC adapter to an outlet controlled by a switch to ensure power is supplied to unit at all times.
NOTE– Do not connect the DC power cord to an outlet that is not constantly energized.
NOTE– A fully charged battery, on the 7305P series will provide approximately 60 minutes of continuous operation at a zero vacuum level (free flow). NOTE– If unit is not in use for extended periods, battery should be recharged every 6 months minimum.
CAUTION– Discharging the battery completely will shorten the life of the battery. Do not operate the unit more than a few minutes if the low battery indica- tor light is lit. Recharge as soon as possible.
NOTE– When charging the battery, the battery charging indicator light will remain lit. If your unit does not hold a charge, please be sure the model you are using has a battery installed prior to returning it to your authorized Sunrise Medical provider or to Sunrise Medical. Verify that the charge light turns on when the external power is applied with the switch in the “Off” position.
NOTE– Units equipped with an internal rechargeable battery contain a sealed
CLEANING INSTRUCTIONS
Collection Bottle:
1.Shut off unit using power switch and allow vacuum to drop. Disconnect power source from the DC input receptacle on the unit.
2.With the collection bottle still in the holder, remove the lid. The bottle can now be taken out of the holder to be emptied.
NOTE– Collection bottle should be emptied and cleaned after each use. The 800ml disposable collection bottle and lid are meant for
3.Collection bottles and lids should be thoroughly cleaned after each use by one of the following methods:
All Bottles:
lWash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry.
1,200ml Reusable Bottle only:
lWash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then wash with rubbing alcohol and air dry.
lWash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then wash with a commercial
disinfectant, follow disinfectant manufacturer's recommended instructions and dilution rates carefully.
lClean the 1200 ml autoclavable jar, lid assembly, gasket, seal washer, float guide, and float thoroughly with mild soap. Rinse thoroughly with
clean water and dry thoroughly. Place parts in autoclave as follows ensuring the parts are not touching: l Jar with open end down.