A&D HW-200KGL Accumulation Function, Operation and Switches, Function table Meaning & Purpose

Page 27

10 Accumulation Function

This function counts the number of weighed items, calculates the total mass value and can display the number and accumulated mass value.

It is necessary to set the parameters of the “accumulation function (f8 )” in the function table in advance to use this function.

It is necessary to set the parameters of the “print mode ( f9 )” in the function table in advance to use the optional built-in printer.

Operation and Switches

The display of the accumulation count has a blinking M+ or

without a unit.

The display of the accumulation value has a blinking M+ kg or

kg with a unit.

Pressing the

MODE

switch, the accumulation count and accumulation value is displayed.

Pressing the

ZERO

switch in the accumulation function while a blinking M+ or is

displayed, resets the current function. The count and accumulated value become zero.

When the optional built-in printer is installed and the PRINT switch is pressed, the accumulation data, date and data number are printed. Date is set at function table f16 .

Caution

The accumulation function can only be used with the first weighing unit accumulated, e.g. you cannot accumulate pcs while displaying kg.

Parameter List and Word Definition

The “near-zero” is within±4 digits from the zero point in the unit of kg. The “digit”, a unit of display, is equivalent to the minimum graduation. The “zero point” is the fundamental starting point to weighing.

Function table

Meaning & Purpose

f8

0

Accumulation function not used.

 

 

The scale accumulates the data, if the F switch is pressed when

 

 

the display is a positive stable value above near-zero. The next

f8

1

accumulation can be performed after the display returns to

 

 

near-zero or to a negative value.

f8

2

The scale accumulates the data, if the F switch is pressed, when

 

 

the display is a stable value and outside near-zero. The next

 

 

accumulation can be performed after the display returns to

 

 

near-zero.

f8

3

When the display is a positive stable value, the scale accumulates

 

 

the data automatically. The next accumulation can be performed

 

 

after the display returns to near-zero or to a negative value.

 

 

 

HW-G Instruction Manual Page 25

HW-G OZIM v1a

Image 27
Contents HW-G Series Page Contents HW-G Ozim v1a Compliance with FCC rules ComplianceOutline and Features Unpacking All AccessoriesType Products Accessories Accessories and Options listOptions List Order code or option name AccessoriesPrecautions for Installing the Scale Precautions for Operating the ScalePrecautions for Storing the Scale Fit the Pan Installing the ScaleSecure the Column Stpe 3 Secure the Display UnitAdjust the Display Unit Connect the PowerLevel the Scale Installing the Batteries for Type L Names OFF Display and SymbolsHW-G Ozim v1a Switches HW-G Ozim v1a Basic Operation Turning the Scale On/Off & Basic WeighingMemo Type L with Batteries Semi-Automatic Tare Input by Weighing Tare and Net DisplayMode Switch Changing Unit & Mode Storing a Unit Mass Counting ModeCounting the number of articles Storing a 100% Mass Percentage ModeReading Percentage Operation and Switches Accumulation FunctionFunction table Meaning & Purpose Preparation Setting Parameters Example Operation and Performance ExamplesFunction table Meaning and Purpose Comparator FunctionComparator Sign Comparison ConditionSwitch to display an item Step If one Has Been selected, press Case Full/Dribble Batch Function Comparator Full/Dribble batch sub-functionHold Parameter and display an item of the full/dribble Switch to display the HI Function Table Meaning and purpose Simple batch function Simple Batch FunctionPreparation Setting Parameters Mode Calibration Adjusting the Scale Gravity Acceleration Table Gravity Acceleration Correction Complete Calibration ProcedureCalibration of the Zero Point Function Table Procedure for Setting Parameters15.2 Parameter List Function or full/dribble batch function Display Meaning and purposePrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Format Data FormatFormat Preparation and Performance Examples Stream ModeCommand Mode Cancel of tareTare Upper limit value Preset tareFinal value Preparation Setting Parameters Pin connections Pin No Signal name Direction Description Options17.1 RS-232C / Relay output / Buzzer OP-03 17.2 RS-422 / RS-485 / Relay output OP-04 Pin No Signal name Direction DescriptionInstalling OP-04 Communication Format Lower limit value ZeroFinal value Reset of Accumulation data and numberAccumulation data output Accumulation count outputSpecification Internal Printer for Type V OP-06Consumables Installation DisassemblyPerformance Check HW-GV HW-G seriesDimensions HW-G Ozim v1a This situation Confirm these items MaintenanceCheck Points Before Calling Maintenance Repair