11.1 Preparation (Setting Parameters)
Step 1 Turn off the display.
Press the ON/OFF switch while the ZERO switch is pressed and held. The function table is displayed.
Step | 2 | Press the | ENTER | switch to display an item of the | |
|
| accumulation function ( f6 ). | |||
Step | 3 | Select a parameter of the comparator function | |||
|
| ( f6 0 ~ | f6 7 ) with the ^ switch. | ||
Step | 4 | Press the |
|
| switch to store the new parameter. |
| ENTER |
Step 5 Press the F switch and the ENTER switch to exit from the function table. Then the scale displays the weighing mode.