A&D HW-100KGL, HG-60KGV, HW-60KGL, HW-200KGL Printing condition when using option OP-06 for type

Page 48

 

Item

Display

Meaning and purpose

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accumulation

 

 

When the display becomes a stable value, the scale

 

 

 

 

accumulates the data automatically. Next accumulation can

 

 

 

 

occur after the display becomes near-zero.

 

 

f8

4

Use: Recording number and mass of articles removed from

 

 

 

 

the pan. (Put articles on the pan. Press

TARE

switch at each

 

 

 

 

removal.)

 

 

 

 

At each finish of full/dribble batch function, the scale

 

 

 

 

accumulates the data automatically.

 

 

f8

5

Use: Packing articles like a powder, it is used for recording

 

 

 

 

the bag number and total mass.

 

 

 

 

The printing condition when using option OP-06 for type V.

 

 

f9

0 #

No print.

 

 

 

f9

1

Pressing the

 

 

switch, the data is output.

 

 

PRINT

 

 

 

 

Auto-print +

 

 

 

 

When the display becomes a positive stable value above

 

 

f9

2

near-zero, the scale outputs the data automatically. Next

 

 

 

 

output can be performed after the display becomes near-

 

 

 

 

zero or negative value.

 

Print Mode

 

 

Auto-print +/-

 

 

 

 

When the display becomes a stable value outside near-zero,

 

 

f9

3

the scale outputs the data automatically. Next output can be

 

 

 

 

performed after the display becomes near-zero.

 

 

f9

4

At each finish of full/dribble batch function, the data is output.

 

 

f9

5

When the accumulation function is used, the data is output.

 

 

 

 

The details of the full/dribble batch function (B$￿')

 

 

f10 0 #

Reaching final value and pressing the F switch, the current

 

 

 

 

process is finished.

 

 

f10 1

Reaching final value and displaying the stability mark, the

 

 

 

 

current process is finished automatically.

 

 

f10 2

Pressing the

 

 

switch, the scale automatically tares and

 

 

 

SET

 

 

 

 

starts the full/dribble batch process. Reaching final value and

 

 

 

 

pressing the F switch, the current process is finished.

 

Full/Dribble

f10 3

Pressing the

 

 

switch, the scale automatically tares and

 

 

SET

 

Batch

 

 

starts the full/dribble batch process. Reaching final value and

 

sub-function

 

 

displaying the stability mark, the current process is finished.

Address for RS-422 / RS-485 for option OP-04

Address

f11

00

# Use This address can be set from 01 to 99. Allows a

to

 

computer to control a scale with the assigned address.

f11

99

(RS-232C should be set to “00”.)

HW-G OZIM v1a

HW-G Instruction Manual Page 46

Image 48
Contents HW-G Series Page Contents HW-G Ozim v1a Compliance Compliance with FCC rulesOutline and Features All Accessories UnpackingAccessories and Options list Type Products AccessoriesOrder code or option name Accessories Options ListPrecautions for Installing the Scale Precautions for Operating the ScalePrecautions for Storing the Scale Installing the Scale Secure the ColumnStpe 3 Secure the Display Unit Fit the PanAdjust the Display Unit Connect the PowerLevel the Scale Installing the Batteries for Type L Names Display and Symbols OFFHW-G Ozim v1a Switches HW-G Ozim v1a Basic Operation Turning the Scale On/Off & Basic WeighingMemo Type L with Batteries Tare and Net Display Semi-Automatic Tare Input by WeighingMode Switch Changing Unit & Mode Counting Mode Storing a Unit MassCounting the number of articles Percentage Mode Storing a 100% MassReading Percentage Operation and Switches Accumulation FunctionFunction table Meaning & Purpose Preparation Setting Parameters Operation and Performance Examples ExampleComparator Function Comparator SignComparison Condition Function table Meaning and PurposeSwitch to display an item Step If one Has Been selected, press Case Full/Dribble Batch Function Comparator Full/Dribble batch sub-functionHold Parameter and display an item of the full/dribble Switch to display the HI Simple Batch Function Function Table Meaning and purpose Simple batch functionPreparation Setting Parameters Mode Calibration Adjusting the Scale Gravity Acceleration Table Complete Calibration Procedure Gravity Acceleration CorrectionCalibration of the Zero Point Procedure for Setting Parameters Function Table15.2 Parameter List Display Meaning and purpose Function or full/dribble batch functionPrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Data Format FormatFormat Stream Mode Preparation and Performance ExamplesCommand Mode Cancel of tareTare Preset tare Upper limit valueFinal value Preparation Setting Parameters Pin connections Pin No Signal name Direction Description Options17.1 RS-232C / Relay output / Buzzer OP-03 Pin No Signal name Direction Description 17.2 RS-422 / RS-485 / Relay output OP-04Installing OP-04 Communication Format Zero Lower limit valueReset of Accumulation data and number Accumulation data outputAccumulation count output Final valueSpecification Internal Printer for Type V OP-06Consumables Disassembly InstallationPerformance Check HW-G series HW-GVDimensions HW-G Ozim v1a Maintenance Check Points Before Calling MaintenanceRepair This situation Confirm these items