14.2 The Complete Calibration Procedure

14.2.1 Gravity Acceleration Correction

Step1

Turn on the display.

 

Open the rear cover of the display unit.

 

Inside the display unit, press and hold the

 

 

switch to enter the calibration mode.

 

CAL

 

Then

Cal 0

is displayed.

Step2

Press the

 

 

 

switch to enter the gravity

ZERO

 

acceleration

 

correction mode.

Step3

Set your local gravity acceleration using

 

the following switches.

 

^ switch

Selecting the value of a digit.

 

< switch

Selecting a digit.

Step4

Press the

 

 

 

 

switch to store the new

ENTER

 

value.

 

 

 

 

 

Step5

Press the

 

 

switch again. Then the

CAL

 

scale returns to the normal weighing

 

mode.

 

 

 

 

 

14.2.2 Preparation

Step

6

Confirm the environmental conditions as follows:

 

 

Maintain a constant temperature and a stable power supply.

 

 

Install the scale on a solid floor or bench where there is no draft,

 

 

vibration, strong magnetic fields or direct sunlight.

 

 

Consider section “4. Caution”.

Step

7

Display normal weighing for at least 30 minutes to warm up the scale.

HW-G Instruction Manual Page 41

HW-G OZIM v1a

Image 43
Contents HW-G Series Page Contents HW-G Ozim v1a Compliance with FCC rules ComplianceOutline and Features Unpacking All AccessoriesType Products Accessories Accessories and Options listOptions List Order code or option name AccessoriesPrecautions for Operating the Scale Precautions for Installing the ScalePrecautions for Storing the Scale Fit the Pan Installing the ScaleSecure the Column Stpe 3 Secure the Display UnitConnect the Power Adjust the Display UnitLevel the Scale Installing the Batteries for Type L Names OFF Display and SymbolsHW-G Ozim v1a Switches HW-G Ozim v1a Turning the Scale On/Off & Basic Weighing Basic OperationMemo Type L with Batteries Semi-Automatic Tare Input by Weighing Tare and Net DisplayMode Switch Changing Unit & Mode Storing a Unit Mass Counting ModeCounting the number of articles Storing a 100% Mass Percentage ModeReading Percentage Accumulation Function Operation and SwitchesFunction table Meaning & Purpose Preparation Setting Parameters Example Operation and Performance ExamplesFunction table Meaning and Purpose Comparator FunctionComparator Sign Comparison ConditionSwitch to display an item Step If one Has Been selected, press Case Full/Dribble Batch Function Full/Dribble batch sub-function ComparatorHold Parameter and display an item of the full/dribble Switch to display the HI Function Table Meaning and purpose Simple batch function Simple Batch FunctionPreparation Setting Parameters Mode Calibration Adjusting the Scale Gravity Acceleration Table Gravity Acceleration Correction Complete Calibration ProcedureCalibration of the Zero Point Function Table Procedure for Setting Parameters15.2 Parameter List Function or full/dribble batch function Display Meaning and purposePrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Format Data FormatFormat Preparation and Performance Examples Stream ModeCancel of tare Command ModeTare Upper limit value Preset tareFinal value Preparation Setting Parameters Options Pin connections Pin No Signal name Direction Description17.1 RS-232C / Relay output / Buzzer OP-03 17.2 RS-422 / RS-485 / Relay output OP-04 Pin No Signal name Direction DescriptionInstalling OP-04 Communication Format Lower limit value ZeroFinal value Reset of Accumulation data and numberAccumulation data output Accumulation count outputInternal Printer for Type V OP-06 SpecificationConsumables Installation DisassemblyPerformance Check HW-GV HW-G seriesDimensions HW-G Ozim v1a This situation Confirm these items MaintenanceCheck Points Before Calling Maintenance Repair