- INSTALLATION MANUAL -

TAB.5

FAULT

CAUSE

REMEDY

No voltage on

No power supply

Restore power supply

Anomalous stop

Blown fuse

Replace fuses

 

Loosened connections

Check connection fitting

Compressor failure

High and Low-pressure pressure-

Ask for skilled assistance

 

switch on

Ask for skilled assistance

 

Clicker on

Ask for skilled assistance

 

Contactor failure

Ask for skilled assistance

 

Compressor thermal relay on

 

The compressor is working but

Frosted evaporator

Open the door and carry out defrost cycle

the cabinet is not cooling

No coolant inside the refrigerating

Ask for skilled assistance

 

system

Ask for skilled assistance

 

Delivery solenoid valve failure

 

Evaporator fans are not working

Fan failure or short-circuit

Ask for skilled assistance

 

Door micro failure

Ask for skilled assistance

The cycle cannot start

Wrong cycle programming

Check time and temperature parameters

AL High Press

High Pressure Switch

Qualified technician required

AL Room Probe

Room Probe interrupted

Qualified technician required

AL Evap Probe

Evap Probe interrupted

Qualified technician required

AL Cond Probe

Cond Probe interrupted

Qualified technician required

AL Needle Probe

Needle Probe interrupted

If a probe cycle is running, the cycle is time-controlled

 

 

Qualified technician required

AL Insert Needle

Needle Probe not inserted

Check the probe inserting cone

High T Room

Room Temp above set value

If the temperature is not within the specified range, apply

 

 

to a qualified technician

Low T Room

Room Temp below set value

If the temperature is not within the specified range, apply

 

 

to a qualified technician

AL BlackOut

No power supply

When power is restored, check the max. temperature

 

 

reached inside the room

AL Door Open

QC room door open

Close the door

 

Door micro faulty

Qualified technician required