Agilent Technologies E1564A, E1563A TRIGgerIMMediate, TRIGgerLEVel, Subsystem Syntax TRIGger

Models: E1563A E1564A

1 173
Download 173 pages 42.87 Kb
Page 104
Image 104

TRIGger

The TRIGger command subsystem controls the behavior of the trigger system.

Subsystem Syntax TRIGger

[:IMMediate]

:LEVel <channel> <level> MIN MAX :LEVel <channel>?

:MODE NORMal MASTer0,2,4,6 SLAVe0,2,4,6 :MODE?

:SLOPe[<n>] POS 1 NEG 0 :SLOPe[<n>]?

:SOURce[<n>] OFF BUS EXT HOLD IMMediate INTernal1-4 TTLT0-7

:SOURce[<n>]?

TRIGger[:IMMediate]

TRIGger[:IMMediate] causes the instrument to transition to the wait-for-sample state immediately, regardless of the trigger source selected.

Comments Instrument Must be Initiated: The instrument must be initiated (INITiate command) and be in the wait-for-trigger state when TRIG:IMM is executed. A “Trigger ignored” error will be generated if the instrument has not been initiated prior to this command or if it is not in the wait-for-trigger state.

Executable when initiated: YES

Coupled command: NO

Reset (*RST) condition: None

TRIGger:LEVel

TRIGger:LEVel<channel> <voltage> MIN MAX sets the level on the specified channel that can be used for internally triggering the instrument. This command is valid only for TRIGger:SOURce INTernal1-4.

Parameters

Name

Type

Range of Values

Default Value

 

 

 

 

voltage

numeric

see Comments

volts

 

 

 

 

104 Digitizers Command Reference

Chapter 3

Page 104
Image 104
Agilent Technologies E1564A, E1563A user manual TRIGgerIMMediate, TRIGgerLEVel, Subsystem Syntax TRIGger