Agilent Technologies E1564A, E1563A user manual CALibrationSOURce?, CALibrationSTATe?

Models: E1563A E1564A

1 173
Download 173 pages 42.87 Kb
Page 58
Image 58

Executable when initiated: NO

Coupled Command: NO

Reset (*RST) Condition: EXTernal

CALibration:SOURce?

CALibration:SOURce? queries which calibration source is set. This setting is shared by all channels. Returns “INT” for INTernal or “EXT” for EXTernal.

CALibration:STATe

CALibration:STATe ON 1 OFF 0 enables the calibration of the instrument. Many instrument operations are not allowed when this state is ON and will result in an error “Illegal while calibrating”. You must set the calibration state to OFF when calibration is finished.

NOTE Sending CAL:STAT OFF, without storing any modified cal constants with the CAL:STORe command, will generate an error. Send the ABORt or *RST command to abort a calibration without storing cal constants.

Executable when initiated: NO

Coupled Command: NO

Reset (*RST) Condition: OFF

CALibration:STATe?

CALibration:STATe? queries the present calibration state of the instrument. A return value of “1” indicates the instrument is enabled and will accept calibration commands and perform calibrations. A return value of “0” indicates the instrument is not calibration enabled and attempting to execute a calibration process command such as CAL:GAIN or CAL:ZERO, will return the error “Calibration not enabled”.

58 Digitizers Command Reference

Chapter 3

Page 58
Image 58
Agilent Technologies E1564A, E1563A user manual CALibrationSOURce?, CALibrationSTATe?