Front Panel Indicators
Failed LED: Illuminates momentarily during digitizer
Access LED: Illuminates when the backplane is communicating with the digitizer.
Error LED: Illuminates only when an error is present in the digitizer’s driver error queue. The error can result from improperly executing a command or the digitizer being unable to pass
Sample LED: Illuminates while the digitizer samples the input for a measurement. Typically blinks for slow sample rates and is on
User Input Terminals
The E1563A Digitizer front panel contains two female connectors for user inputs. Mating male connectors are supplied with the module. However, the user must provide the input cable and connect the male connector to the cable. See "User Cabling Considerations" for recommended
External Trigger Input
The front panel contains a
plug&play
Figure 1-1. E1563A 2-Channel Digitizer Front Panel