Appendix A

Digitizers Specifications

General

Number of channels:

 

 

 

Selectable input filters (2-pole Bessel):

 

E1563A: 2 channels

 

 

 

E1563A (per channel): 25 kHz

 

E1564A: 4 channels

 

 

 

E1564A (per channel): 1.5kHz, 6 kHz, 25 kHz, 100 kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

Timing:

 

 

 

 

 

 

Memory/Triggering:

 

 

 

Bandwidth:>400 kHz for all ranges

 

 

Trigger: Time and Event

 

 

 

Resolution: 14 bits (including sign)

 

 

Pre-trigger capture: Yes

 

 

 

Sample rates: 1 Sa/s to 800 kSa/s

 

 

Memory: 4 Mbyte to 128 Mbyte PC SIMM

 

Integral Non-linearity (all ranges): 2.5 LSB

FIFO memory

 

 

 

 

Built-in DSP: No

 

 

 

Minimum External Trigger Pulse Width: ~ 20 nsec

Alias protection: Oversample

 

 

External Sample Latency: 100 nsec (due to optocoupler)

Time base resolution:

0.1 msec

 

 

External Trigger Latency: ±0.5 x Sample Interval

Low-frequency CMRR: 113 dB

 

 

Minimum Ext Sample Clock Pulse Width: ~ 20 nsec

 

 

 

 

 

 

 

 

Can accept non-periodic sample pulses

 

 

 

 

 

 

 

 

 

 

 

 

Cooling/Slot:

 

 

 

Environmental:

 

 

 

 

Watts/slot

 

 

 

 

 

 

For indoor use, pollution degree 2 (IEC 61010-1)

 

E1563A:

20.6W

 

 

 

Operating altitude: 3000 meters or mainframe altitude

E1564A:

37.4W

 

 

 

specification, whichever is lower

 

DP mm H2O:0.18

 

 

 

Operating temperature: 0°C to 55°C

 

Air flow liter/s: 2.8

 

 

 

Rel humidity: up to 80% at 31°C, decreasing to 50% at 40°C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E1563A/E1564A Accuracy Specifications (1 Year)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Range

 

 

Zero Offset1

Zero Offset1

 

Gain

 

Noise

 

 

 

(with filter OFF)

(with filter ON)

(% of reading)

 

(3 sigma)

 

 

 

 

 

 

 

 

 

 

 

 

 

Specifi-

 

Temperature

Specifi-

 

Temperature

Specifi-

Temperature

 

Specifi-

 

 

cation2

 

Coefficient3

cation2

 

Coefficient3

cation2

Coefficient3

 

cation

 

 

 

 

 

 

 

 

 

 

 

 

 

0.0625V

 

 

20 mV

 

1.9 mV/°C

28 mV

 

4.3 mV/°C

0.034%

0.0061%/°C

 

57 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

0.25V

 

 

78 mV

 

6 mV/°C

110 mV

 

16 mV/°C

0.034%

0.0061%/°C

 

180 mV

 

 

 

 

 

 

 

 

 

 

 

 

1V

 

300 mV

 

15 mV/°C

430 mV

 

63 mV/°C

0.034%

0.0061%/°C

 

720 mV

 

 

 

 

 

 

 

 

 

 

 

 

4V

 

1.2 mV

 

60 mV/°C

1.7 mV

 

251 mV/°C

0.034%

0.0061%/°C

 

2.88 mV

 

 

 

 

 

 

 

 

 

 

 

 

16V

 

21 mV

 

1.3 mV/°C

21 mV

 

1.63 mV/°C

0.034%

0.0061%/°C

 

14.7 mV

 

 

 

 

 

 

 

 

 

 

 

 

64V

 

28 mV

 

1.65 mV/°C

34 mV

 

4.24 mV/°C

0.034%

0.0061%/°C

 

48 mV

 

 

 

 

 

 

 

 

 

 

 

 

256V

 

79 mV

 

4.28 mV/°C

110 mV

 

16.2 mV/°C

0.034%

0.0061%/°C

 

189 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

1Valid within the range of 0°C to 55°C. A zero offset calibration for all channels must be performed if the instrument experiences a temperature <0°C or >55°C for these specifications to remain valid.

2Specification is valid when tested at a temperature within ±5°C of the calibration temperature.

3 Amount of error that must be added for each °C outside of ±5°C of the calibration temperature.

Appendix A

Digitizers Specifications 119

Page 119
Image 119
Agilent Technologies E1563A Appendix a Digitizers Specifications, Timing Memory/Triggering, Cooling/Slot Environmental

E1563A, E1564A specifications

Agilent Technologies, a prominent player in the field of electronic measurement, offers a range of products designed for various testing applications. Among their notable offerings are the E1564A and E1563A modules. Both models are designed for high-performance testing and measurement, catering particularly to the needs of engineers and technicians in the telecommunications and wireless industries.

The Agilent E1564A is a high-speed data and synchronization generation module, specifically tailored for advanced signal analysis. One of its primary features is its ability to provide a wide range of modulation types, including QPSK and 16QAM, which are essential for testing modern communication systems. The module also boasts a flexible output that supports multiple channels, making it efficient for testing complex systems where multiple data streams are present. Its exceptional performance in generating accurate waveforms enables engineers to perform thorough tests and validations on their devices.

On the other hand, the E1563A module is predominantly focused on protocol analysis and provides engineers with insights into the behavior of their communication systems. This module features comprehensive support for various communication protocols, making it invaluable when debugging and validating designs. Engineers can utilize the extensive measurement capabilities of the E1563A to analyze data integrity and system performance, ensuring that their products meet required specifications.

Both modules leverage advanced technologies from Agilent, ensuring high accuracy and reliability in measurements. They are equipped with powerful processing capabilities that allow for real-time data analysis, a critical aspect when working with high-frequency signals. Furthermore, these modules are designed for seamless integration into Agilent’s test and measurement platforms, enhancing usability and facilitating complex testing scenarios.

In summary, the Agilent E1564A and E1563A modules represent cutting-edge solutions for engineers involved in testing and validating communication systems. With their robust features, including support for various modulation schemes and protocol analysis, these modules enable efficient and accurate measurements. The commitment to quality and precision from Agilent Technologies continues to make these devices essential tools in the industry, helping engineers push the boundaries of innovation in telecommunications.