Comments Changing Ranges: The present range setting will determine the maximum and minimum values that can be entered without error. Changing range will keep the level at the same percentage of the new range. For example, if level is set to 2.0 on the 4V range, the level is set to 8.0 if you change to the 16V range (50% of full range).

Setting Levels: Changing ranges will change an existing level to the same percent of full scale on the new range. For example, if an 8.0 level is set on the 16V range and the range is then changed to the 4V range, the level attempts to change to 2.0V (still 50%). However, for this range, this action causes an error message to be generated and the new level is set to the maximum or minimum the new range will support.

Trigger Slopes: TRIG:SLOPe specifies the direction of signal movement through which the level will trigger the digitizer. TRIG:SLOPe POSitive causes a trigger when the signal passes through the level and rises above the specified level. TRIG:SLOPe NEGative causes a trigger when the signal passes through the level and falls below the specified level.

Executable when initiated: NO

Coupled command: YES. Range setting

Reset (*RST) condition: 0.00 on all channels

TRIGger:LEVel?

TRIGger:LEVel<channel>? queries the value of the trigger level set on the specified channel.

TRIGger:MODE

Parameters

Comments

TRIGger:MODE NORMal MASTer<n> SLAVe<n> sets the trigger mode. Master and Slave parameters set the modules for use in connecting more than one module together for simultaneous measurements from the same trigger and sample.

Name

Type

Range of Values

Default Value

 

 

 

 

<n>

numeric

0, 2, 4, 6

none

 

 

 

 

Master and Slave Operation: NORMal sets standard trigger operation and the specified trigger and sample sources are used. MASTer<n> and SLAVe<n> pairs a sample line and a trigger line which are then used for multiple unit synchronization. See Chapter 2 for more information, including diagrams.

Chapter 3

Digitizers Command Reference 105

Page 105
Image 105
Agilent Technologies E1563A, E1564A user manual TRIGgerLEVel?, TRIGgerMODE, Reset *RST condition 0.00 on all channels

E1563A, E1564A specifications

Agilent Technologies, a prominent player in the field of electronic measurement, offers a range of products designed for various testing applications. Among their notable offerings are the E1564A and E1563A modules. Both models are designed for high-performance testing and measurement, catering particularly to the needs of engineers and technicians in the telecommunications and wireless industries.

The Agilent E1564A is a high-speed data and synchronization generation module, specifically tailored for advanced signal analysis. One of its primary features is its ability to provide a wide range of modulation types, including QPSK and 16QAM, which are essential for testing modern communication systems. The module also boasts a flexible output that supports multiple channels, making it efficient for testing complex systems where multiple data streams are present. Its exceptional performance in generating accurate waveforms enables engineers to perform thorough tests and validations on their devices.

On the other hand, the E1563A module is predominantly focused on protocol analysis and provides engineers with insights into the behavior of their communication systems. This module features comprehensive support for various communication protocols, making it invaluable when debugging and validating designs. Engineers can utilize the extensive measurement capabilities of the E1563A to analyze data integrity and system performance, ensuring that their products meet required specifications.

Both modules leverage advanced technologies from Agilent, ensuring high accuracy and reliability in measurements. They are equipped with powerful processing capabilities that allow for real-time data analysis, a critical aspect when working with high-frequency signals. Furthermore, these modules are designed for seamless integration into Agilent’s test and measurement platforms, enhancing usability and facilitating complex testing scenarios.

In summary, the Agilent E1564A and E1563A modules represent cutting-edge solutions for engineers involved in testing and validating communication systems. With their robust features, including support for various modulation schemes and protocol analysis, these modules enable efficient and accurate measurements. The commitment to quality and precision from Agilent Technologies continues to make these devices essential tools in the industry, helping engineers push the boundaries of innovation in telecommunications.