Executable when initiated: YES

Coupled command: NO

Related commands: *OPC?, *WAI

Reset (*RST) condition: none

*OPC?

*OPC? causes the E1563A and E1564A Digitizers to wait for all pending operations to complete. A single ASCII “1” is then placed in the output queue.

Comments INIT vs. OPC?: The INIT command is considered complete when the measurement is started. *OPC? will return “1” once INIT is processed and measurements start but the instrument may not be finished taking all readings initiated.

Executable when initiated: YES

Coupled command: NO

Related commands: *OPC?, *WAI

Reset (*RST) condition: none

*RST

*RST resets the E1563A and E1564A Digitizers as follows:

Sets all commands to their *RST state

Aborts a calibration (CAL:STATe ON)

Resets the CAL:STATe to OFF

Aborts all pending operations.

*RST does not affect:

The output queue

The Service Request and Standard Event Status Enable Registers

The enable unmasks for the QUEStionable Status Registers

Calibration data

Comments Executable when initiated: Yes

Coupled command: No

Reset (*RST) condition: none

112 Digitizers Command Reference

Chapter 3

Page 112
Image 112