Executable when initiated: YES
Coupled command: NO
Related commands: *OPC?, *WAI
Reset (*RST) condition: none
*OPC?
*OPC? causes the E1563A and E1564A Digitizers to wait for all pending operations to complete. A single ASCII “1” is then placed in the output queue.
Comments INIT vs. OPC?: The INIT command is considered complete when the measurement is started. *OPC? will return “1” once INIT is processed and measurements start but the instrument may not be finished taking all readings initiated.
Executable when initiated: YES
Coupled command: NO
Related commands: *OPC?, *WAI
Reset (*RST) condition: none
*RST
*RST resets the E1563A and E1564A Digitizers as follows:
•Sets all commands to their *RST state
•Aborts a calibration (CAL:STATe ON)
•Resets the CAL:STATe to OFF
•Aborts all pending operations.
*RST does not affect:
•The output queue
•The Service Request and Standard Event Status Enable Registers
•The enable unmasks for the QUEStionable Status Registers
•Calibration data
Comments Executable when initiated: Yes
Coupled command: No
Reset (*RST) condition: none