Image 85

[SENSe:]

Subsystem Syntax

[SENSe:]DATA?

The SENSe command subsystem is used to change low-level parameters such as voltage range, sweep and sweep offset points and to set the reference oscillator source and frequency. It is also used to obtain measurement data from the module.

[SENSe:]

DATA? <rdgs_per_channel>[,channel_list]

DATA:ALL? <rdgs_per_channel>

DATA:COUNt?

DATA:CVTable? [channel_list]

ROSCillator:EXTernal:FREQuency <freq>l

ROSCillator:EXTernal:FREQuency?

ROSCillator:SOURce INTernal EXTernal

ROSCillator:SOURce?

SWEep:OFFSet:POINts <neg_value> MIN MAX

SWEep:OFFSet:POINts? MIN MAX

SWEep:POINts <neg_value> MIN MAX

SWEep:POINts? MIN MAX

VOLTage[<channel>][:DC]:RANGe <range> MIN MAX

VOLTage[<channel>][:DC]:RANGe?

VOLTage[<channel>][:DC]:RESolution?

[SENSe:]DATA? <rdgs_per_channel>[,channel_list] returns voltage formatted data from all channels (default) or only from the specified channel list. <channel_list> has the form (@1) or (@2), (@1,2), (@1:4) or (@1,2,3,4). For specific channels, but not all, the format is (@1,3,4).

Parameters

Name

Type

Range of Values

Default Value

 

 

 

 

rdgs_per_

numeric

1 to MAX samples

none

channel

 

depends on size of RAM

 

 

 

on module

 

 

 

(see SAMPle:COUNt)

 

 

 

 

 

channel_list

numeric

1-2 (E1563A)

N/A

 

 

1-4 (E1564A)

 

 

 

 

 

Chapter 3